Recall Z-2031-2012

Becton Dickinson & Co. · initiated April 12, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD Gram Stain Kit, containing 1- 250 mL bottle of Gram Crystal Violet, and additional reagents: Gram Iodine (Stabilized), Gram Decolorizer, Gram Safranin. Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.

Codes / lots: Lot 1348873, Exp April 30, 2013

Reason for recall

Potential for increase level of artifacts to appear as large, spherical and easy to distinguish or small & more difficult to distinguish from gram positive cocci.

FDA-determined cause: Material/Component Contamination

Action taken

BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard the affected product. An Acknowledgement Form was attached for customers to complete and return via fax to 410-316-4258. Contact the BD Customer Service Department for questions regarding this recall.

Details

Recall number
Z-2031-2012
Initiated
April 12, 2012
Posted by FDA
July 18, 2012
Terminated
December 10, 2012
Product code
JTS – Stains, Microbiologic
Quantity
539 kits
Distribution
Nationwide Distribution.
Recalling firm
Becton Dickinson & Co., Sparks, MD

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2032-2012 Class IIGram Crystal Violet 250 mL, catalog #212525, in cartons of 4 bottles. Gram Stain Kits…Potential for increase level of artifacts to appear as large, spherical and easy to distinguish or small &… 04/12/2012Terminated