Recall Z-2046-2009

Terumo Cardiovascular Systems Corporation · initiated February 5, 2008

Terminated

Product

Terumo Sarns cardiovascular disposable products, Weighted Flexible Sucker, sterile, Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; Catalog No. 804113.

Codes / lots: Lots 0394979, 0400254, 0403180, 0405486, 0411500, 0414942, 0420022, 0424246, 0427742, 0430190, 0435074, 0436862, 0443077, 0452056, 0465817, 0470928, 0479748, 0484668 and 0491029.

Reason for recall

A loose strand of stainless steel left on the device during manufacturing may result in an unretrieved metal fragment remaining in the patient.

FDA-determined cause: Process control

Action taken

Consignees were notified via Urgent - Medical Device Removal letters dated 2/5/08 to cease use of the product. Terumo CVS will contact customers to arrange replacement for the affected product upon receipt of the attached response form. Questions are to be directed to Terumo CVS Customer Service at 1-800-521-2818.

Details

Recall number
Z-2046-2009
Initiated
February 5, 2008
Posted by FDA
August 4, 2009
Terminated
November 1, 2009
Product code
DTS – Sucker, Cardiotomy Return, Cardiopulmonary Bypass
Quantity
7079
Distribution
Worldwide Distribution --- including USA and countries of Belgium, Brazil, Dominican Republic, Germany, Jordan and Latin America.
Recalling firm
Terumo Cardiovascular Systems Corporation, Ann Arbor, MI

Official FDA recall record