Recall Z-2048-2020

Becton Dickinson & Co. · initiated April 14, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD BBLTM Gram Stain Kit with Stabilized Iodine-- IVD For the differential staining of bacteria, Catalog Number: 212539 - Product Usage: Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.

Codes / lots: Lot Number/ Exp Date: 0007261 06/30/21; 0007255 05/31/21; 0028309 03/30/21. Expanded 5/12/20-Lot Numbers/Exp. Date: 0041114 07/31/21; 0014322 06/30/21

Reason for recall

May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain

FDA-determined cause: Under Investigation by firm

Action taken

BD issued Urgent Medical Device Notification to Customers/Distributors on 4/14/20 via FedEX stating reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific Catalog and lot numbers listed above. Destroy all product subject to the recall following your institutions process for destruction. 2. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form. Even if you do not have any of the affected lots in your inventory, please complete the Customer Recall Response Form indicating you have zero (0) quantity and return as indicated above. Distributors requested to provide customer list to BD. BD will provide a product replacement for all discarded inventory. BD Customer/Technical Support 800-638-8663, Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside the United States, contact your local BDrepresentative or distributor. On 5/12/20 firm expanded the recall to an additional six (6) lots of product.

Details

Recall number
Z-2048-2020
Initiated
April 14, 2020
Terminated
February 10, 2022
Product code
JTS – Stains, Microbiologic
Quantity
1469 kits
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Taiwan, Hong Kong, Chile, EMEA, Thailand, China, Japan, Australia, New Zealand, Korea, Costa Rica, Mexico, Canada, and Singapore.
Recalling firm
Becton Dickinson & Co., Sparks, MD

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2046-2020 Class IIBD BBLTM Gram Crystal Violet 250mL-- IVD Primary stain for Staining Microorganisms…May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci… 04/14/2020Terminated
Z-2047-2020 Class IIBD BBLTM Gram Crystal Violet 3.8L-- IVD Primary stain for Staining Microorganisms…May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci… 04/14/2020Terminated