Recall Z-2060-2018

Shimadzu Medical Systems · initiated January 12, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intended to be used for the radiography in the hospital with X-ray devices. This device is operated and used by the physicians and X-ray technologist. The object of this device is total patient population.

Codes / lots: Serial #: 0161G68101, 0261G65303, 0261G65702, 0261G68204, 3M72BFB21001, 3M72BFB2A001, 3M72BFC2A001, 3M72BFC2C001, 3M72BFB32001, 3M72BFB32006, 3M72BFB33003,3M72BFC33001, 3M72BFC34001, 3M72BFB35005, 3M72BFC37002, 3M72BFC42001, 3M72BFB46002

Reason for recall

When selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray Generator Console, as well as the "integral dose" on the external console of fluoroscopy, were reset to 0. There has been one report of this event.

FDA-determined cause: Software design

Action taken

Shimadf0u sent an Urgent Voluntary Medical Device Recall Notice letter dated January 2018. The letter identified the affected product, problem and actions to be taken. For questions contact Shimadzu representative or contact Shimadzu National Technical Support at 800-228-1429.

Details

Recall number
Z-2060-2018
Initiated
January 12, 2018
Terminated
February 24, 2022
Product code
IZI – System, X-Ray, Angiographic
Quantity
17
Distribution
US Nationwide in the states of IL, WA, MI, TX, AZ, CA, LA, FL, SC, and NJ.
Recalling firm
Shimadzu Medical Systems, Torrance, CA

Official FDA recall record