Recall Z-2074-2020

Dako Denmark A/S · initiated November 5, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and Autostainer Plus, Model S3800.

Codes / lots: All serial numbers of devices manufactured between 7/22/1997-6/28/2018 that do not have an installed syringe tray, UDI 05700572035497.

Reason for recall

The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm issued letters dated 10/30/2019 via email or standard mail, or phone call beginning 11/5/2019 notifying the customer of the issue, providing the risk to health, and actions that needed to be taken. On 09/03/2020, Agilent Dako sent Urgent Field Safety Notices to customers who were asked to take the following actions: 1) Carefully read and communicate to those within your organization of the updates stated in this notification letter and attached Autostainer Link Basic User Guide (rev. 06). Your local sales team will confirm with you that you have read and acknowledged this notification. 2) Your Agilent representative should be contacted immediately for assistance in the event of leakage. 3) Please discard previous versions of the Autostainer Link Basic User Guide that you are currently using.

Details

Recall number
Z-2074-2020
Initiated
November 5, 2019
Posted by FDA
May 21, 2020
Terminated
May 17, 2022
Product code
KPA – Slide Stainer, Automated
Quantity
796 devices
Distribution
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand…
Recalling firm
Dako Denmark A/S, Glostrup