Recall Z-2079-2012
Getinge USA Inc · initiated May 25, 2012
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128
Codes / lots: Serial numbers: W50024258; W50024259; W50026573; W50025790; W50017381; W50020296; W50028138; W50020952
Reason for recall
Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error.
FDA-determined cause: Nonconforming Material/Component
Action taken
Getinge USA issued a Device Correction Notice dated June 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to perform a trip test to determine if they have a functioning ground fault breaker. Customers will be contacted by a Getinge representative to schedule an appointment to exchange the earth fault breaker. Customers were provided a Customer Response Form to complete and fax back to 585-272-5033. For questions customers should call 800-475-9040, ext. 5270. For questions regarding this recall call 585-475-1400.
Details
- Recall number
- Z-2079-2012
- Initiated
- May 25, 2012
- Posted by FDA
- July 26, 2012
- Terminated
- May 18, 2016
- Product code
- MEC – Disinfector, Medical Devices
- Quantity
- 8 washers
- Distribution
- Worldwide Distribution - USA including Maryland, Indiana, New Mexico, Massachuesetts, Texas, Kansas and the country of Canada
- Recalling firm
- Getinge USA Inc, Rochester, NY