Recall Z-2081-2016

Alcon Research, Ltd. · initiated May 24, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

CENTURION Vision System

Codes / lots: Serial Number 1601444801X

Reason for recall

The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.

FDA-determined cause: Nonconforming Material/Component

Action taken

Alcon has already notified the affected customer verbally on May 24, 2016, and via written correspondence on June 1, 2016. The device was corrected on May 26, 2016, and the Power Entry Switch that was removed from the customers CENTURION Vision System did not exhibit a grounding pin susceptible to detachment. For further questions, please call (817) 293-0450.

Details

Recall number
Z-2081-2016
Initiated
May 24, 2016
Posted by FDA
June 28, 2016
Terminated
April 13, 2017
Product code
HQC – Unit, Phacofragmentation
Quantity
1 unit
Distribution
US Distribution to Illinois
Recalling firm
Alcon Research, Ltd., Fort Worth, TX

Official FDA recall record