Recall Z-2082-2021

Fresenius Medical Care Holdings, Inc. · initiated April 28, 2021

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

(1) 08-4078-BB BIBAG 650 GRAMS/BAG 16 BAGS/CASE (2) 08-4080-BB BIBAG 900 grams/bag 12 bags/case For patients with acute and chronic end stage renal disease going under PD in healthcare or home setting GRANUFLO & NATURALYTE: Concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Codes / lots: (1) Lot Number: B4MA27140 BAG GTIN CASE GTIN BIBAG 650 GRAMS 08-4078-BB 00840861101313 10840861101310 (2) Lot Number: B5MB24110 BAG GTIN CASE GTIN BIBAG 900 GRAMS 08-4080-BB 00840861101320 10840861101327

Reason for recall

Exposed to freezing temperatures due to refrigerated truck malfunctions

FDA-determined cause: Environmental control

Action taken

Fresenius issued telephone script communication on April 28.2021 stating reason for recall, heath risk and action to take: Tell them you already contacted their Home Therapy Nurse to alert them that you would be calling them. o The purpose of my call today is that you received a delivery from us on (either Feb 22nd or Feb 23rd). o The truck you received a delivery from had a temperature malfunction due to the bad storms Texas experienced back in February. This caused the product you were delivered to be exposed to freezing or near freezing temperatures for about 3 hours. o o I see you received on . o (if received any PD SOLUTION/SALINE/or CONCENTRATES): Explain that the solutions inside their product are not affected and that we have freeze/thaw studies that prove this. Explain that due to the freezing temperatures, the packaging may become compromised, and they may have seen leaks. For the other products such as Hand Sanitizer and other ancillary products, again, the product could be compromised according to their packaging requirements listed on their packaging. o We are giving you the opportunity to return these products if you choose to do so at this time. o If returning product(s): o Fill out RGA spreadsheet with customer information including how much product they are returning (include lot# and amount returning). Driver will pick up returned products on their next scheduled visit. Fresenius Corrections and Removals Rep: Email Post Market Surveillance to open a complaint against product. o If not returning product(s): o Thank them for their time and reiterate to always inspect their packaging for leaks as this is what they are trained to do anyways from their Home Therapies Nurse. If you have any questions call 781-699-2243 or email: beth.st.germain@fmc-na.com

Details

Recall number
Z-2082-2021
Initiated
April 28, 2021
Terminated
August 8, 2022
Product code
KPO – Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
Quantity
(1) 20 cases (2) 30 cases
Distribution
US Distribution to: TX only
Recalling firm
Fresenius Medical Care Holdings, Inc., Waltham, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2083-2021 Class IIThe Drain Bag Set (tubings and flexible bags) is a passive, closed drainage system used…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2084-2021 Class IICitrasate Dry Acid Concentrate and GranuFlo Dry Acid Concentrate are each intended to be…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2085-2021 Class II026-20036A Liberty Drain Line, Luer Tap For patients with acute and chronic end stage…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2086-2021 Class IICitrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2087-2021 Class IINaturaLyte Dry Bicarbonate Concentrate is composed of sodium bicarbonate powder and is…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2088-2021 Class IIThe Fresenius APD Luer-Lock Adapter, Catalog number 050-95018, is designed to connect a…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2089-2021 Class IIThe MTS Set is a single-use, sterile (EO), non-pyrogenic, optional accessory that…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated
Z-2090-2021 Class II050-95012 Sterile Stay Safe Cap The Indications for Use/Intended Use for the Fresenius…Exposed to freezing temperatures due to refrigerated truck malfunctions 04/28/2021Terminated