Recall Z-2084-2014

DePuy Orthopaedics, Inc. · initiated June 3, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical Instrument, used to ensure the locking taper surface of the Distal Stem is protected during proximal preparation.

Codes / lots: Product # 197614140; Lot # 107310, 107311, 107313, 107314, 107315, 136244, 136247, 136251, 136252, 136256, 136257, 136262, 136268, 136271, 136276, 136286, 196221, 196222, 196225, 225871, 225872, 225873, 228185, 228187, 228188, 235886, 235887, 240634, 240635, 246147, 246179, 254812, 254814, 260373, 260374, 264629, 264630, 281910, 281911, 281912, 281915, 288356, 288359, 288361, 288363, 442821, 451586, 452549, 453409, 455523, 455524, 456513, 456514, 456515, 457650 & 457651. Product # 197614190; Lot # 290958, 290959, 290960, 310940, 310941, 310942, 310947, 310948, 327468, 327469, 327475, 327477, 327478, 327480, 365792, 365793, 365794, 365795, 365796 & 365797. Product # 197615140; Lot # 107320, 107323, 107325, 107328, 107330, 107331, 136369, 136370, 136371, 136372, 136373, 136374, 136375, 136376, 136377, 136377A, 136378, 136380, 136381, 196227, 196228, 225874, 225875, 225876, 228196, 228197, 228198, 231202, 235888, 235890, 239254, 239255, 251499, 251500, 251502, 251504, 254817, 254818, 260375, 260377, 264633, 264638, 383749, 393633, 405087, 420828, 421141, 432535, 442823, 451846, 457652, 465564, 465565, 480737, 482786, 491057, 497256, 503931, 518456, 523799 & 526931. Product #…

Reason for recall

The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.

FDA-determined cause: Device Design

Action taken

DePuy Synthes sent an Urgent Information - Device Correction Notice letter dated June 24, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: 1. Immediate Remedy: Discard the RECLAIM Taper Sleeve Protector instrument before stem implantation. 2. Changes to the literature and IFU: Documents are revised as follows: " RECLAIM Revision Hip Systems Surgical Technique # DSUS/JRC/0614/0200 pages 11, 12, 13, 16, 30, 34, 35 (figure numbering from pages 11 to 27 were also revised): Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems Surgical Technique FAQ # DSUS/JRC/0614/0202 pages 1 and 2: Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems IFU #090200799 REV E Page 5: Verbiage will be revised and implemented as soon as possible, depending on regulatory approvals. 3. Change: The company consulted with design surgeons and developed a strategy to eliminate the taper sleeve protector from the system. As a result of removing the taper sleeve protector, the company is evaluating whether there is a need for a protective component to be used during the distribution Customers with questions were instructed to call 574-371-4917. For questions regarding this recall call 574-371-4577.

Details

Recall number
Z-2084-2014
Initiated
June 3, 2014
Posted by FDA
July 21, 2014
Terminated
May 17, 2017
Product code
KWA – Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Quantity
3,623
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to AUSTRALIA, AUSTRIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IBERIA, IRELAND, ISRAEL, ITALY, NEW ZEALAND, NORWAY…
Recalling firm
DePuy Orthopaedics, Inc., Warsaw, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2082-2014 Class IIRECLAIM Taper Protector Sleeve. RECLAIM Revision Hip System Orthopedic Manual Surgical…The product can be difficult to remove from the Distal Stem both out of the package and after proximal… 06/03/2014Terminated
Z-2083-2014 Class IIRECLAIM DISTAL TAPERED, Angled Distal Stem, various sizes. Taper Protector Sleeve.…The product can be difficult to remove from the Distal Stem both out of the package and after proximal… 06/03/2014Terminated