Recall Z-2084-2017

Beckman Coulter Inc. · initiated April 3, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Salicylate Reagent (SALY 2 x 45), Catalog No. 378194. Intended for quantitative determination of salicylate concentration in human serum or plasma.

Codes / lots: all lots

Reason for recall

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

FDA-determined cause: Under Investigation by firm

Action taken

An Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. Customers are informed of the action and resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center -Via http://www.beckmancoulter.com/customersupport/support -Via phone, call 1-800-854-3633 in the United States and Canada Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.

Details

Recall number
Z-2084-2017
Initiated
April 3, 2017
Terminated
May 31, 2018
Product code
DKJ – Colorimetry, Salicylate
Quantity
50,645 units total (43,816 units in US)
Distribution
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar…
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2078-2017 Class IILactate Reagent (LACT 2 x 50), Catalog No. A95550 Intended for the quantitative…Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole… 04/03/2017Terminated
Z-2079-2017 Class IIUric Acid Reagent (URIC 2 x 300), Catalog No. 442785. Intended for the quantitative…Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole… 04/03/2017Terminated
Z-2080-2017 Class IIEnzymatic Creatinine Reagent (CR-E 2 x 200), Catalog No. A60298. Intended for the…Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole… 04/03/2017Terminated
Z-2081-2017 Class IICholesterol Reagent(CHOL 2 x 300), Catalog No. 467825 Intended for quantitative…Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole… 04/03/2017Terminated
Z-2082-2017 Class IIHDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative…Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole… 04/03/2017Terminated
Z-2083-2017 Class IITriglycerides Reagent (TG 2 x 300), Catalog No. 445850. Intended for quantitative…Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole… 04/03/2017Terminated