Recall Z-2089-2014

Perouse Medical · initiated May 10, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter.

Codes / lots: Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448.

Reason for recall

Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.

FDA-determined cause: Process control

Action taken

Perouse Medical initiated this recall by sending an e-mail notification to consignees on May 10, 2012. A formal written recall notification letter was sent to consignees on May 16, 2012. The recall letter, written in English, on company's letterhead, dated May 16, 2012, titled "VOLUNTARY BATCH RECALL Of FLAMINGO Inflations Devices", notified the consignee of the following: product information with codes and batch codes, product description, description of the potential default, reason for the recall action, instructions to the consignee, contact information and assistance. Response form "INVENTORY FORM".

Details

Recall number
Z-2089-2014
Initiated
May 10, 2012
Posted by FDA
July 22, 2014
Terminated
September 9, 2014
Product code
MAV – Syringe, Balloon Inflation
Quantity
2,496 units
Distribution
US.
Recalling firm
Perouse Medical, Irigny