Recall Z-2089-2015

Eizo Corporation · initiated July 7, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

RadiForce RX850, 8MP Color LCD Monitor, Part Code No. 0FTD1958A2, UPC 690592037180. This product is intended to be used in displaying and viewing digital images those of digital mammography for review and analysis by trained medical practitioners.

Codes / lots: 220045044, 20050044

Reason for recall

EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular graphics boards with DisplayPort.

FDA-determined cause: Device Design

Action taken

EIZO sent An Urgent Field Safety Notice dated July 7, 2015, to all customers who purchased the RadiForce RX440, RX650, RX850 and GX540. The notice informs customers that the RadiForce monitors could display abornormally in 6-bit (64 gradations) instead of normal 8-bit (256 gradations) or 10-bit (1024 gradations) while connecting to particular graphics boards with DisplayPort. The notice provides the customers with the actions to be taken and contact information (1-800-800-5202, M-F, 8am-4pm, PST, us.feedback@eizo.com, 5710 Warland Drive, Cypress, CA 90630).

Details

Recall number
Z-2089-2015
Initiated
July 7, 2015
Posted by FDA
July 16, 2015
Terminated
November 30, 2016
Product code
PGY – Display, Diagnostic Radiology
Quantity
2 units
Distribution
Nationwide Distribution including OH, FL, IL, TX, OH, NY, AZ, and NJ.
Recalling firm
Eizo Corporation

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2087-2015 Class IIRadiForce GX540, 5MP Monochrome LCD Monitor, Part Code No. 0FTD1826A5, 0FTD1826B5…EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular… 07/07/2015Terminated
Z-2088-2015 Class IIRadiForce RX650, 6MP Color LCD Monitor, Part Code No. 0FTD1957A5, 0FTD195B5, UPC…EIZO is recalling the RadiForce LCD Monitors could display abnormally while connecting to particular… 07/07/2015Terminated