Recall Z-2089-2025

Siemens Healthcare Diagnostics, Inc. · initiated June 4, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ADVIA 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)

Codes / lots: Siemens Material Number 10318905; UDI: 00630414473413, 00630414473413; Lot No. TP253035, TP253045.

Reason for recall

Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.

FDA-determined cause: Under Investigation by firm

Action taken

An Urgent Medical Device Correction notification dated 6/4/25 was mailed to consignees. This notification provides consignees with calibration instructions given the issue found in affected lots. Additionally, instructions are given on how to proceed with product depending on performance. Consignees are to discontinue use of and discard of the calibrator SETPOINT lots impacted by this recall. Consignee locations are to review the provided notification with their Medical Directors to determine the appropriate course of action. A copy of the recall notification should be retained in laboratory records and shared with users. Consignees are to complete and return the provided Effectiveness Check Form within 30 days of receipt of the recall notification. Consignees with any questions can contact their Siemens Customer Care Center or local Siemens Healthineers technical support representative.

Details

Recall number
Z-2089-2025
Initiated
June 4, 2025
Posted by FDA
July 3, 2025
Product code
JCN – Control, Cell Counter, Normal And Abnormal
Quantity
3,266 units
Distribution
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Tarrytown, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2086-2025 Class IIADVIA 120/2120/2120i SETpoint CalibratorProducts have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems. 06/04/2025Open, Classified
Z-2087-2025 Class IIADVIA 120/2120/2120i 3-in-1 TESTpoint Control (ABN1 Control)Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems. 06/04/2025Open, Classified
Z-2088-2025 Class IIADVIA 120/2120/2120i 3-in-1 TESTpoint Control (Norm Control)Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems. 06/04/2025Open, Classified