Recall Z-2093-2016

Smith & Nephew, Inc. · initiated May 26, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

LEGION(TM) HK FEMORAL ASSEMBLY, SIZE 4 RIGHT, A/P 58.5 MM, M/L 66 MM, REF 71421364 The LEGION Hinge Knee System is indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight and activity level are compatible with an adequate long-term result. 3. Failed osteotomies, unicompartmental replacement, or total knee replacement. 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, 5. Constrained and hinge knee systems are designed for use in patients in primary and revision surgery. The LEGION Hinge Knee System is for cemented use only.

Codes / lots: Lot codes: 15KM20807,15KM20808,15KM20811,15KM20812,15LM08258,15LM12482,15MM03005,15MM03007

Reason for recall

The surface inside of the femoral component, where cement is applied for adhesion, is below specification.

FDA-determined cause: Process control

Action taken

The firm sent an Urgent Product Recall letter dated May 26, 2016 to customers, via email and FedEx Overnight delivery . The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. Inspect inventory and locate any unused devices from the listed product and batch numbers, and quarantine them immediately. 2. Complete the Inventory Return Certification Form on the following page, indicating the quantities that need to be returned. If you do not have product to return, place an x in the column No Product to Return. Include your contact information in the spaces provided at the bottom of the form. 3. Contact Smith & Nephews Field Actions Department via email at FieldActions@smith-nephew.com or fax +1-901-566-7975 to return the completed Inventory Return Certification Form and obtain a return authorization (RA) number if you have product to return. 4. Return any affected product to the address listed on the Inventory Return Certification Form. Indicate the RA number on your return shipment.

Details

Recall number
Z-2093-2016
Initiated
May 26, 2016
Terminated
August 16, 2019
Product code
KRO – Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
Quantity
16 units
Distribution
Worldwide Distribution and US Nationwide including Puerto Rico and the countries of Australia, Austria, Belgium, Canada, France, Germany, Italy, Netherlands, South Africa, Sweden, and UK.
Recalling firm
Smith & Nephew, Inc., Memphis, TN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2090-2016 Class IILEGION(TM) HK FEMORAL ASSEMBLY, SIZE 3 LEFT, A/P 54.5 MM, M/L 62 MM, REF 71421373 The…The surface inside of the femoral component, where cement is applied for adhesion, is below specification. 05/26/2016Terminated
Z-2091-2016 Class IILEGION(TM) HK FEMORAL ASSEMBLY, SIZE 3 RIGHT, A/P 54.5 MM, M/L 62 MM, REF 71421363 The…The surface inside of the femoral component, where cement is applied for adhesion, is below specification. 05/26/2016Terminated
Z-2092-2016 Class IILEGION(TM) HK FEMORAL ASSEMBLY, SIZE 4 LEFT, A/P 58.5 MM, M/L 66 MM, REF 71421374 The…The surface inside of the femoral component, where cement is applied for adhesion, is below specification. 05/26/2016Terminated
Z-2094-2016 Class IILEGION(TM) HK FEMORAL ASSEMBLY, SIZE 5 LEFT, A/P 62 MM, M/L 70 MM, REF 71421375 The…The surface inside of the femoral component, where cement is applied for adhesion, is below specification. 05/26/2016Terminated
Z-2095-2016 Class IILEGION(TM) HK FEMORAL ASSEMBLY, SIZE 5 RIGHT, A/P 62 MM, M/L 70 MM, REF 71421365 The…The surface inside of the femoral component, where cement is applied for adhesion, is below specification. 05/26/2016Terminated