Recall Z-2110-2023

Abbott · initiated June 12, 2023

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac).

Codes / lots: UDI/DI 05415067036025, Batch Numbers: 8597228, 8598983, 8601676, 8611443, 8619197, 8627877, 8650955, 8663679, 8795668, 8815423, 8824607, 8830530, 8846835

Reason for recall

There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used.

FDA-determined cause: Device Design

Action taken

Abbott issued a MEDICAL DEVICE RECALL notice to its consignees beginning on 06/12/2023. The notice explained the problem with the device, the risk, and requested the return of all units. Should you have questions about this issue, please contact your local Abbott Representative or Abbott Support at 1-800-544-1664 (Option 2) (U.S.), 7:00 a.m. - 7:00 p.m. Central Time, Monday through Friday.

Details

Recall number
Z-2110-2023
Initiated
June 12, 2023
Posted by FDA
July 11, 2023
Product code
DQY – Catheter, Percutaneous
Quantity
675 units
Distribution
Worldwide distribution - US Nationwide and the countries of Slovakia, Czech Republic.
Recalling firm
Abbott, Plymouth, MN

Official FDA recall record