Recall Z-2130-2020

Teleflex Medical Inc · initiated October 11, 2019

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Sprotte Needle, REF 33115131B w/ Intro 20 G x 3.5", distributed by Teleflex Medical The SPROTTE lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/discography procedures.

Codes / lots: Lot/Batch Numbers: 74A1900703, 74D1802540, 74E1800978, 74F1801413, 74G1800896, 74H1800162, 74K1801738, 74K1802909, 74L1800502

Reason for recall

Teleflex received the attached Medical Device Recall Notice from Pajunk Medical. Teleflex Medical is a distributor of Pajunk products. According to the notice, Pajunk internally identified a packaging problem that may affect certain batches of the SPROTTE lumbar puncture cannulas with introducer. Due to this problem Pajunk cannot guarantee the sterility of the affected batches.

FDA-determined cause: Process control

Action taken

The firm, Teleflex Medical, initiated the recall by letter on 10/11/2019. The notice requested that the consignees take actions as follows: Distributors: 1. Immediately discontinue distribution and quarantine any products with the product codes and lot numbers listed in the letter. 2. Using the provided customer letter and recall acknowledgement form templates, communicate this recall to any of your customers who have received product included within the scope of the recall. 3. Have the customers return any affected product to you, together with a completed Recall Acknowledgement Form, for consolidation and return to Teleflex Medical. In the event that an alternative approach is needed, contact Teleflex Customer Service for more information at 1-866-396-2111. 4. To return recalled products to Teleflex Medical, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical. 5. Once you have completed returning all of the recalled products from your own inventory, and collecting and consolidating all of the recalled products from your customers, please check the box on the enclosed Recall Acknowledgment Form that indicates that you have completed the recall and fax it to 1-855-419-8507, Attn: Customer Service, or email it to recalls@teleflex.com. This will allow us to document completion of the recall. 6. If you and your customers have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter. Customer of Teleflex Products: 1. If you have affected stock in inventory, immed

Details

Recall number
Z-2130-2020
Initiated
October 11, 2019
Product code
BSP – Needle, Conduction, Anesthetic (W/Wo Introducer)
Quantity
800 devices
Distribution
Nationwide distribution.
Recalling firm
Teleflex Medical Inc, Morrisville, NC

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2129-2020 Class IISprotte Needle, REF 32115130C w/ Intro 22 G x 3.5", distributed by Teleflex Medical The…Teleflex received the attached Medical Device Recall Notice from Pajunk Medical. Teleflex Medical is a… 10/11/2019Open, Classified