Recall Z-2131-2015

Volcano Corporation · initiated June 22, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5. Radiology: The Volcano s5/s5i and CORE Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature

Codes / lots: Model s5. Part numbers: 804200-002 and 804200-014

Reason for recall

A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain circumstances.

FDA-determined cause: Software design

Action taken

Urgent Medical Device Correction letters were sent on June 22, 2015 to all affected sites via FED Ex.

Details

Recall number
Z-2131-2015
Initiated
June 22, 2015
Posted by FDA
July 20, 2015
Terminated
April 14, 2017
Product code
IYO – System, Imaging, Pulsed Echo, Ultrasonic
Quantity
Total of 4007 devices, all models
Distribution
Worldwide Distribution-US (nationwide) and regions in Europe and the countries of Japan, Canada, South Africa, Taiwan, Thailand, Singapore, Philippines, Australia, New Zealand, El Salvador, Malaysia, Mexico, Lebanon, Kuwait, South Korea, Kenya, Jordan, Iran, India, Israel, Hong Kong, Egypt, Algeria, Dominican Republic, Costa Rica, Chile, Brazil, Bolivia, and Argentina
Recalling firm
Volcano Corporation, Rancho Cordova, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2132-2015 Class IIVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with…A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain… 06/22/2015Terminated
Z-2133-2015 Class IIVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with…A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain… 06/22/2015Terminated
Z-2134-2015 Class IIVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with…A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain… 06/22/2015Terminated
Z-2135-2015 Class IIVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with…A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain… 06/22/2015Terminated
Z-2136-2015 Class IIVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with…A software issue was discovered where an inaccurate FFR/iFR value could be calculated under certain… 06/22/2015Terminated