Recall Z-2132-2020

Smith & Nephew, Inc. · initiated January 28, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5, CATALOG NUMBER: REF 74035825 SIZE 6, CATALOG NUMBER: REF 74035826 SIZE 7, CATALOG NUMBER: REF 74035827 SIZE 8, CATALOG NUMBER: REF 74035828 SIZE 9, CATALOG NUMBER: REF 74035829 SIZE 10, CATALOG NUMBER: REF 74035830

Codes / lots: SIZE 1, CATALOG NUMBER: REF 74035821; LOT/BATCH CODES: 19EM19112R, 19FM05587R, 19FM11174R, 19HM18075, 19HM19648, 19HM21015, 19HM21016, 19HM23242 SIZE 2, CATALOG NUMBER: REF 74035822; LOT/BATCH CODES: 19EM21573R, 19EM21573S, 19FM05589R, 19HM21031, 19HM21033, 19HM21878, 19HM21881 SIZE 3, CATALOG NUMBER: REF 74035823; LOT/BATCH CODES: 19EM21578R, 19HM19009, 19JM09476, 19JM18509, 19JM18525, 19JM21072, 19KM14473, 19KM17780 SIZE 4, CATALOG NUMBER: REF 74035824; LOT/BATCH CODES: 19EM14365R, 19EM14365S, 19FM11173, 19HM18078, 19HM19659, 19HM21861, 19JM21075, 19KM15573, 19KM1557 SIZE 5, CATALOG NUMBER: REF 74035825; LOT/BATCH CODES: 19EM14367R, 19HM18079, 19HM21855, 19KM12683 SIZE 6, CATALOG NUMBER: REF 74035826; LOT/BATCH CODES: 19EM24480R, 19HM18080, 19HM21856, 19KM21512 SIZE 7, CATALOG NUMBER: REF 74035827; LOT/BATCH CODES: 19EM14371R, 19FM01063R, 19FM11177R, 19HM18335, 19HM21873, 19HM21874, 19KM17781 SIZE 8, CATALOG NUMBER: REF 74035828; LOT/BATCH CODES: 19EM24498R, 19EM24498S, 19FM05591R, 19FM11178R, 19FM11178S, 19JM07464, 19JM07465, 19KM07278, 19KM12699 SIZE 9, CATALOG NUMBER: REF 74035829; LOT/BATCH CODES: 19EM21582R, 19FM11179R, 19GM21787, 19JM07467, 19KM07276, 19KM15575, 19KM15576…

Reason for recall

The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.

FDA-determined cause: Process design

Action taken

The firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.

Details

Recall number
Z-2132-2020
Initiated
January 28, 2020
Terminated
December 11, 2023
Product code
HSX – Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
Quantity
200 units
Distribution
US Nationwide distribution.
Recalling firm
Smith & Nephew, Inc., Memphis, TN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2133-2020 Class IIJOURNEY II UNI RESECTION PREP TRIAL, RIGHT MEDIAL/ LEFT LATERAL - Product Usage: The…The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to… 01/28/2020Terminated