Recall Z-2143-2026
Medline Industries, LP · initiated March 24, 2026
Class I – dangerous or defective products that could cause serious health problems or death
Product
Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. ANGIOGRAPHY CATH LAB, Medline Kit Number/SKUDYNJ23532C; 2. CATH LAB PACK, Medline Kit Number/SKUDYNJ65185; 3. ANGIO KIT, Medline Kit Number/SKUVASC1057.
Codes / lots: Medline Kit Number/SKU DYNJ23532C: UDI/DI each 10193489460872, UDI/DI case 40193489460873, Lot Number: 25DMA809; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23GBA252; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23HBI873; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23IBR043; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 23KBF277; Medline Kit Number/SKU DYNJ65185: UDI/DI each 10193489253146, UDI/DI case 40193489253147, Lot Number: 24ABQ959; Medline Kit Number/SKU VASC1057: UDI/DI each 10889942431922, UDI/DI case 40889942431923, Lot Number: 25KMF194.
Reason for recall
Medline has identified the presence of particulate within the fluid path of the Manifolds.
FDA-determined cause: Under Investigation by firm
Action taken
Medline Industries, LP issued an URGENT MEDICAL DEVICE notice to its consignees on 3/24/2026 via email and mail. The notice explained the issue, potential risk and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-046 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Details
- Recall number
- Z-2143-2026
- Initiated
- March 24, 2026
- Posted by FDA
- May 8, 2026
- Product code
- OEQ – Angiography/Angioplasty Kit
- Quantity
- 508 kits
- Distribution
- US (nationwide) and AE, AU, CA, JP, KR, NL, SG, SK.
- Recalling firm
- Medline Industries, LP, Northfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2138-2026 Class I | Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2139-2026 Class I | Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038301…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2140-2026 Class I | Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038402…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2141-2026 Class I | Namic Preceptor Manifold, custom, Medline Product Number/SKU 64037107; Extravascular…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2142-2026 Class I | Namic Preceptor Manifold, custom, Medline Product Number/SKU 64039307; Extravascular…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2144-2026 Class I | 1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS-LEFT HEART,KIT, Medline…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2145-2026 Class I | 1. 3 VALVE MAN KIT, Medline Kit SKU Number: 65101544; 2. 3V MANIFOLD KIT, Medline Kit…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2146-2026 Class I | Medline medical convenience kits, containing Namic Manifold, Labeled as: HEART CATH…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2147-2026 Class I | Medline medical convenience kits, containing Namic Manifold, Labeled as: 1.…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |
| Z-2148-2026 Class I | Medline medical convenience kits, containing Namic Manifold, Labeled as: TAVI, Medline…Medline has identified the presence of particulate within the fluid path of the Manifolds. | 03/24/2026Open, Classified |