Recall Z-2146-2010

Baxa Corporation · initiated May 26, 2010

Terminated

Product

Exacta-Mix 2400 Valve Assembly w/Calibration Bag, Sterile R, 250 mL, Order No. REF: 724, Baxa Corporation.

Codes / lots: All lot numbers

Reason for recall

Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid pathway.

FDA-determined cause: Process design

Action taken

Consignees were notified by e-mail and Baxa Bag Product Notice letter on 05/26/2010 with instructions to not use any bags if the fill port cap is detached. Photos of a detached cap and properly attached cap were included. Defective bags could be returned to Baxa for credit. Information available at 800-678-2292.

Details

Recall number
Z-2146-2010
Initiated
May 26, 2010
Posted by FDA
August 2, 2010
Terminated
August 4, 2011
Product code
KPE – Container, I.V.
Quantity
410,050 units
Distribution
Worldwide distribution, including USA, Argentina, Australia, Bermuda, Brazil, Canada, China, Colombia, Hong Kong, Malaysia, Mexico, New Zealand, Pakistan, Panama, South Korea, Taiwan, United Kingdom, and Uruguay.
Recalling firm
Baxa Corporation, Englewood, CO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2137-2010 Exacta-Mix EVA Container, Sterile R, 250 mL, Order No. REF: 137, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2138-2010 Exacta-Mix EVA Container, Sterile R, 500 mL, Order No. REF: 138, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2139-2010 Exacta-Mix EVA Container, Sterile R, 1000 mL, Order No. REF: 139, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2140-2010 Exacta-Mix EVA Container, Sterile R, 2000 mL, Order No. REF: 140, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2141-2010 Exacta-Mix EVA Container, Sterile R, 3000 mL, Order No. REF: 141, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2142-2010 Exacta-Mix EVA Container, Sterile R, 4000 mL, Order No. REF: 142, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2143-2010 Exacta-Mix EVA Container, Sterile R, 5000 mL, Order No. REF: 143, Baxa Corporation.Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2144-2010 Halobag EVA Dual-Chamber Bag, Sterile R, 3000 mL Total Volume W/500 mL upper chamber…Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2145-2010 Exacta-Mix Calibration Bag, Sterile R, non-pyrogenic fluid path, 250 mL Reorder No. H938…Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2147-2010 Exacta-Mix Valve Set, Six Inlet, Non-Vented, Assembled, Tubing Kit w/Calibration Bag…Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated
Z-2148-2010 Exacta - Mix Valve Set, Six Inlet, Tubing Kit w/Calibration Bag, Fluid Path is sterile…Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid… 05/26/2010Terminated