Recall Z-2148-2014

Synthes, Inc. · initiated April 16, 2014

Class I Terminated

Class I – dangerous or defective products that could cause serious health problems or death

Product

DePuy Synthes Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies)

Codes / lots: Part nos. 04.315.003, 04.315.004, 04.315.005, 04.315.006 04.315.023 04.315.024 04.315.025 04.315.026 04.315.027 04.315.028 04.315.053 04.315.054 04.315.055 04.315.056 04.315.063 04.315.064 04.315.065 04.315.066 04.315.067 04.315.068, with lot nos.: 6245205, 7129328, IS10400 6246203, 6582199, 6794535, 6814674, 6877737, 6921359, 6981567, 7008430, 7081952, 7129329, 7460419, 7556148 6246984, 6342006, 6964974, 7129330, 7460420, 6246983, 7389334, 7422735, 7515818 6245460, 6342005, 6454138, 6495399, 6651393, 6816023, 6898959, 6921360, 7306561, 7515824 6307904, 6365443, 6410544, 6438375, 6625839, 6651392, 6984928, 7241620, 7351855, 7390470, 7408179, 7458272, 7458273, 7458274, IS10394 6450702, 6454139, 6553925, 6625840, 6883404, 6921586, 7031344, 7185897, 7389329, 7408181, 7422706, 7422707, 7422708, 7458299, 7515832, IS10385 6393118, 6393119, 6410545, 6454140, 6512859, 6625841, 6651391, 6984925, 7185895, 7408182, 7422703, 7422704, 7422705, 7515829, 7515830 6393120, 7408180, 7515833 7122918, 7389333, 7422700, 7545381, IS10388 6252759, 7129331, 7155891, 7418771, 7448753 6245463, 6438376, 6981568, 6983876,7063885, 7129332,7309797, 7418772, 7418848, 7448754 6251769, 7081953, 7310030, 7365542…

Reason for recall

DePuy Synthes is initiating a recall of certain lots of the Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies) because they may reverse post-operatively.

FDA-determined cause: Device Design

Action taken

The firm, DePuy Synthes, sent "Urgent Notice: Medical Device Recall" letters dated April 16, 2014 to affected customers. The letter identified the affected product, reason for recall, potential patient impact, mitigation steps for patients with distractors currently implanted and actions to be taken. Customers were instructed to review inventory, immediately remove the affected lots from stock, complete verification section of the letter and return the affected product. For questions call 610-719-5450 or contact your DePuy Synthes Sales consultant.

Details

Recall number
Z-2148-2014
Initiated
April 16, 2014
Posted by FDA
August 20, 2014
Terminated
September 9, 2015
Product code
MQN – External Mandibular Fixator And/Or Distractor
Quantity
1499
Distribution
US Nationwide Distribution
Recalling firm
Synthes, Inc., West Chester, PA

Official FDA recall record