Recall Z-2152-2015

Biomet Spine, LLC · initiated March 12, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing Technologies. Spinal fixation component.

Codes / lots: Part Numbers: 14-582020, 14-582022, 14-582025, 14-582030, 14-582035, 14-582040, 14-582120, 14-582122, 14-582125, 14-582127, 14-582130, 14-582135, 14-582140, 14-582220, 14-582222, 14-582225, 14-582230, 14-582235, 14-582240, 14-582245, 14-582250, 14-592320, 14-582325, 14-582330, 14-582335, 14-582340, 14-582345, 14-582350, 14-582530, 14-582535, 14-582540, 14-582545, 14-582550, 14-582565, 14-582570, 14-582575, 14-583020, 14-583025, 14-583030, 14-583035, 14-583040, 14-583120, 14-583125, 14-583130, 14-583135, 14-583140, 14-583220, 14-583225, 14-583230, 14-583235, 14-583240, 14-583245, 14-583250, 14-583320, 14-583325, 14-583330, 14-583335, 14-583340, 14-583345 and 14-583350. All lots.

Reason for recall

Multiaxial screws may not meet internal requirements related to fatigue strength. Compression testing result and surface treatment are out of specification.

FDA-determined cause: Nonconforming Material/Component

Action taken

Biomet Spine issued an Urgent Medical Device Recall Notice/Response Form dated 3/12/2015 to their affected customers.

Details

Recall number
Z-2152-2015
Initiated
March 12, 2015
Posted by FDA
July 20, 2015
Terminated
December 21, 2015
Product code
NKB – Thoracolumbosacral Pedicle Screw System
Quantity
4,765 units
Distribution
Distribution US Nationwide (AL, CA, CO, FL, GA, LA, MA, MD, MI, MO, NY, OH, PA, TX, WA, WI) and The Netherlands.
Recalling firm
Biomet Spine, LLC, Parsippany, NJ

Official FDA recall record