Recall Z-2152-2025
Beckman Coulter Inc. · initiated June 12, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.
Codes / lots: UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C
Reason for recall
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
FDA-determined cause: Nonconforming Material/Component
Action taken
On June 12, 2025, Beckman Coulter issued a "Urgent: Medical Device Recall" notification via Email. On July 8, 2025, Beckman Coulter issued a updated notification to provide additional information on the health risk and actions consignees should take. Beckman Coulter ask consignees to take the following actions: 1. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 2. Please discard any remaining inventory of effective lots per your laboratory protocols and ask for replacement. 2. Please share this information with your laboratory staff and retain this notifications as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 3. Please respond within 10 days in one of the following ways: 1. Electronically, if you received the notification via email, 2. Manually, complete and return the enclosed Response Form.
Details
- Recall number
- Z-2152-2025
- Initiated
- June 12, 2025
- Posted by FDA
- July 21, 2025
- Product code
- CGF – Radioimmunoassay, Estrone
- Quantity
- 496 units
- Distribution
- US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
- Recalling firm
- Beckman Coulter Inc., Brea, CA