Recall Z-2154-2026

GE Medical Systems, LLC · initiated March 26, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20

Codes / lots: UDI-DI: 00840682146616; Serial Numbers: REV2A2200014CN REV2A2200072CN REV2A2100016CN REV2A2200068CN REV2A2200034CN REV2A2200116CN REV2A2200037CN REV2A2000006CN REV2A2200041CN REV2A2100011CN REV2A2100048CN REV2A2200015CN REV2A2100038CN REV2A2200011CN REV242200002CN REV2A2200120CN REV2A2200018CN REV2A2300016CN REV2A2200067CN REV2A2000007CN REV2A2200105CN REV2A2200007CN REV2A2000032CN REV2A2100019CN

Reason for recall

GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

FDA-determined cause: Software design

Action taken

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to recall.23015@gehealthcare.com Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare

Details

Recall number
Z-2154-2026
Initiated
March 26, 2026
Posted by FDA
May 8, 2026
Product code
JAK – System, X-Ray, Tomography, Computed
Quantity
24 units
Distribution
Worldwide - US Nationwide distribution.
Recalling firm
GE Medical Systems, LLC, Waukesha, WI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2149-2026 Class IIRevolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison… 03/26/2026Open, Classified
Z-2150-2026 Class IIRevolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison… 03/26/2026Open, Classified
Z-2151-2026 Class IIRevolution Apex Elite, X-ray Computed Tomography, Model Number 5590000-23GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison… 03/26/2026Open, Classified
Z-2152-2026 Class IIRevolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison… 03/26/2026Open, Classified
Z-2153-2026 Class IIRevolution CT ES, X-ray Computed Tomography, Model Number 5590000-15GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison… 03/26/2026Open, Classified
Z-2155-2026 Class IIRevolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison… 03/26/2026Open, Classified