Recall Z-2155-2012

Stryker Spine · initiated April 23, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-FRANCE; Distributed in the USA by: Stryker Spine. 2 Pearl Ct. Allendale, NJ 07401. Provides access to the thoracic and lumbar spine from a posterior approach via a small incision.

Codes / lots: Lot Code: 093309

Reason for recall

Stryker received three reports involving blades of the Luxor Retractor becoming disassembled from the base causing the retractor to collapse insitu.

FDA-determined cause: Nonconforming Material/Component

Action taken

Stryker sent Urgent Product Recall letters/return response forms dated April 23, 2012 via FedEx to Stryker branches/agencies and hospitals where product might have been distributed. The letter identified the affected product, discussed the issue, potential hazards, risk mitigations, and the return process. Customers are to examine their inventory and reconcile any of the affected product with the information provided. In addition, the Customer Response Form should be completed and faxed back to Regulatory Compliance at 201-760-8370 within 5 days of receipt of the letter. The affected product should also be returned following the instructions provided. Questions should be directed to Dan Roberts, Regulatory Affairs Compliance Specialist at 201-760-8298.

Details

Recall number
Z-2155-2012
Initiated
April 23, 2012
Posted by FDA
August 7, 2012
Terminated
November 13, 2013
Product code
GAD – Retractor
Quantity
30 devices
Distribution
Worldwide Distribution -- US, Switzerland, Netherlands, Brazil, Hong Kong, Germany, and Taiwan.
Recalling firm
Stryker Spine, Allendale, NJ

Official FDA recall record