Recall Z-2159-2017

Mesa Laboratories, Inc. · initiated May 12, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton of 6 of 02.0027). Polyethylene bottle Conductivity/TDS Calibrator solutions are a secondary standard solution used for the calibration of conductivity/TDS cells together with conductivity/TDS measurement instruments. The conductivity/TDS cells and instruments may be indicated for calibrating the conductivity measurement function of hemodialysis machines and water purification equipment for hemodialysis, or for verifying proper function of hemodialysis machine or water purification equipment measurement functions. These solutions are used remotely from the hemodialysis machine or water purification equipment, and do not come into contact with the patient.

Codes / lots: Lot number: ML-C14-1352

Reason for recall

Mesa Laboratories is recalling Conductivity/TDS Calibrator Solution due to failure to meet product specifications.

FDA-determined cause: Process control

Action taken

MesaLabs sent an Device Correction Notice dated May 12, 2017, via certified letter to destroy product by pouring it down the drain and free replacements will be shipped. For further questions, please call (303) 987-8000 ext. 120.

Details

Recall number
Z-2159-2017
Initiated
May 12, 2017
Posted by FDA
May 16, 2017
Terminated
December 19, 2017
Product code
FKH – Solution-Test Standard-Conductivity, Dialysis
Quantity
9900
Distribution
USA (nationwide) Distribution and Internationally to one customer in Canada, and govt/VA/military
Recalling firm
Mesa Laboratories, Inc., Lakewood, CO

Official FDA recall record