Recall Z-2160-2021
GE OEC Medical Systems, Inc · initiated June 10, 2021
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures
Codes / lots: All OEC 3D Systems
Reason for recall
There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
FDA-determined cause: Under Investigation by firm
Action taken
On June 11, 2021, GE Healthcare issued an "Urgent Medical Device Correction via certified mail to all affected consignees. In addition to informing consignees about the recalled product, the firm asked customers to take the following actions: 1. You may continue to use your system. 2. Please disseminate this information to applicable users within your facility. 3. When the system is not in use, it is recommended to connect the system to power with the workstation plugged in and the interconnect cable attached to the mainframe, to reduce potential battery depletion. It is not necessary to turn the workstation power on. 4. Prior to each use, please follow the Start Up Checklist in the System Set Up section (Section 2) of the OEC Elite or OEC 3D Operator Manuals and verify that no error message is displayed on the touch panels or the monitor. a. If the system displays the following error message, as described in the Troubleshooting and messages section (Section 14) of the OEC Elite or OEC 3D Operator Manuals: , this is an indication of a depleted coin cell battery. If this error is displayed, please contact GE Authorized Service. 5. Complete and return the attached response form promptly upon receipt and return to fieldactionssurgery@ge.com. 6. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Details
- Recall number
- Z-2160-2021
- Initiated
- June 10, 2021
- Terminated
- May 9, 2024
- Product code
- OXO – Image-Intensified Fluoroscopic X-Ray System, Mobile
- Quantity
- 3 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore…
- Recalling firm
- GE OEC Medical Systems, Inc, Salt Lake City, UT