Recall Z-2161-2026
Jolife AB · initiated April 13, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 LUCAS 3 Model Number: 99576 Catalog Number: 99576-000043 UDI-DI code: 00883873861875 LUCAS 3.1 Model Number: 99576 Catalog Number: 99576-000063 UDI-DI code: 00883873834329 LUCAS chest compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
Codes / lots: LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 Serial Numbers: 30100846 30112125 30123845 30124348 30124440 30124475 30124991 30136605 30148155 30148365 3015B568 3015B682 3015C276 3015C722 3015C723 3015C724 3015F246 3015F587 3016G845 3016H983 3017K078 3017K079 3017K080 LUCAS 3 Model Number: 99576 Catalog Number: 99576-000043 UDI-DI code: 00883873861875 Serial Numbers: 35160778 35173272 35173461 35185820 35185817 35175040 35175039 35173347 35160921 35175373 35174279 35172701 35161449 LUCAS 3.1 Model Number: 99576 Catalog Number: 99576-000063 UDI-DI code: 00883873834329 Serial Numbers: 3520R763 3522AN95 3519F615 3520N419 3522DK48 3521S574 3521T901 3520R187 3518B300 3522EG06 3520P749 3520R206 3520S005 3520P512 3520P691 3520P741 3520P742 3520P744 3520P745 3520P746 3520P748 3520P810 3520S006 3520P747 3520S003 3521AF21 3518B131 3520L408 3519E743 3521AC18 3521U379 3520O364 3520S292 3519F035 3520S149 3521Z495 3521Z496 3520P082 3520L957 3520R180 3519G235 3519G633 3519H568 3519E607 3522BI32 3518D028 3522CN47 3520K867 3519I425 3520K923 3523FA45 3520O771 3520R123 3521W020 3520K966 3522AV78 3521T568 3520O776 3518A707 3521T325 3521T329…
Reason for recall
Due to a subset of chest compression systems/devices where the service performed on the systems were not appropriately documented (The documents for the service performed to justify the passing results were not include in the service records).
FDA-determined cause: Under Investigation by firm
Action taken
On 04/13/2026, the firm sent via US Postal an "Urgent: Medical device correction" letter informing customers that through complaint reviews, Stryker discovered a subset of work orders performed between July 2023 through November 2025 with Performance Inspection Procedures (PIP) recorded as PASS without the corresponding objective evidence properly documented in the work orders. Customers are instructed to: 1. Immediately check internal inventory to locate the affected product as listed. 2. Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact customers to arrange for corrective action on the device if service has not been performed. 4. There are no additional actions that users can and should take. Continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter Maintaining the Equipment subsection General Maintenance and Testing . 5. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until the correction has been completed. 6. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details, so Stryker can inform the recipients appropriately. Any questions or concerns, contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
Details
- Recall number
- Z-2161-2026
- Initiated
- April 13, 2026
- Posted by FDA
- May 11, 2026
- Product code
- DRM – Compressor, Cardiac, External
- Quantity
- 153 systems
- Distribution
- U.S. Nationwide distribution in the states of AL, AZ, CA, FL, HI, IA, KS, KY, LA, MA, MD, MN, NC, NV, OH, SC, TX, UT, VA, WA, and WY.
- Recalling firm
- Jolife AB, Lund