Target Detachable Coils: TARGET 360 NANO 3.5 MM X 6CM MODEL Number:M0035443560 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Codes / lots: MODEL Number:M0035443560: Lot Number/Expiration Date: 18736425 31-Dec-18 18521457 31-Jul-18 18838703 31-Jan-19 18703233 30-Nov-18 18443288 30-Jun-18 18735209 30-Nov-18 18659056 31-Oct-18 18781251 31-Jan-19 18801665 31-Jan-19 18747459 31-Dec-18 18802980 31-Jan-19 18659087 31-Oct-18 18649087 31-Oct-18 18536525 31-Aug-18 19046694 31-May-19 19047725 31-May-19 18875709 28-Feb-19 18987089 30-Apr-19 19006438 30-Apr-19 18959391 31-Mar-19 18936943 31-Mar-19 19022830 30-Apr-19 18984771 30-Apr-19 18922696 31-Mar-19 18898321 28-Feb-19 19061363 31-May-19 18853805 28-Feb-19.

Reason for recall

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

FDA-determined cause: Process change control

Action taken

Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated June 3, 2016, to all affected customers via FED Ex. Letter includes recall letter, list of impacted lot numbers, customer acknowledgement form and frequently asked questions. Customers were instructed to immediately check their inventory. segregate the effected inventory in a secure location for return to Stryker, circulate the Field Safety Notice internally to all interested/affected parties. Inform Stryker if any of the subject devices have been distributed to other organizations. Customers were also instructed to complete the attached customer response form and return the completed form to their nominated Stryker Representative.

Details

Recall number
Z-2177-2016
Initiated
June 6, 2016
Posted by FDA
July 11, 2016
Terminated
March 3, 2017
Product code
HCG – Device, Neurovascular Embolization
Quantity
45,302 (8,188 in the U.S.) - total, all model numbers
Distribution
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal…
Recalling firm
Stryker Neurovascular, Fremont, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2160-2016 Class IITarget Detachable Coils: TARGET 360 NANO 1 MM X 2 CM MODEL Number: M0035421020…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2161-2016 Class IITarget Detachable Coils: TARGET 360 NANO 1 MM X 3 CM MODEL Number: M0035421030…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2162-2016 Class IITarget Detachable Coils: TARGET 360 NANO 1.5 MM X 2 CM MODEL Number: M0035421520…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2163-2016 Class IITarget Detachable Coils: TARGET 360 NANO 1.5 MM X 3 CM MODEL Number: M0035421530…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2164-2016 Class IITarget Detachable Coils: TARGET 360 NANO 1.5 MM X 4 CM MODEL Number:M0035421540…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2165-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1 MM X 1 CM MODEL Number:M0035431010…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2166-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1 MM X 2 CM MODEL Number:M0035431020…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2167-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1 MM X 3 CM MODEL Number:M0035431030…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2168-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1.5 MM X 1 CM MODEL Number:M0035431510…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2169-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1.5 MM X 2 CM MODEL Number:M0035431520…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2170-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1.5 MM X 3 CM MODEL Number:M0035431530…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2171-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 1.5 MM X 4 CM MODEL Number:M0035431540…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2172-2016 Class IITarget Detachable Coils: TARGET 360 NANO 2MM X 3CM MODEL Number:M0035442030 Neurology:…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2173-2016 Class IITarget Detachable Coils: TARGET 360 NANO 2MM X 4CM MODEL Number:M0035442040 Neurology:…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2174-2016 Class IITarget Detachable Coils: TARGET 360 NANO 2.5 MM X 4CM MODEL Number:M0035442540…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2175-2016 Class IITarget Detachable Coils: TARGET 360 NANO 3MM X 4CM MODEL Number:M0035443040 Neurology:…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2176-2016 Class IITarget Detachable Coils: TARGET 360 NANO 3MM X 6CM MODEL Number:M0035443060 Neurology:…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2178-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 2MM X 2CM MODEL Number:M0035452020…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2179-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 2MM X 3CM MODEL Number:M0035452030…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2180-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 2MM X 4CM MODEL Number:M0035452040…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2181-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 2MM X 6CM MODEL Number:M0035452060…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2182-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 2.5MM X 3CM MODEL Number:M0035452530…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2183-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2184-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035452560…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2185-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035453040…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2186-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 3MM X 6CM MODEL Number:M0035453060…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated
Z-2187-2016 Class IITarget Detachable Coils: TARGET HELICAL NANO 3MM X 8CM MODEL Number:M0035453080…Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended… 06/06/2016Terminated