Recall Z-2182-2018
Siemens Healthcare Diagnostics, Inc. · initiated April 19, 2018
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 17150BF, 17159BD, 17178AC, 17269BD, and 17312BB.
Codes / lots: Lots 17135BC 17150BF 17159BD 17178AC 17269BD 17312BB
Reason for recall
The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators. The negative bias observed for patient, QC, and calibrator samples, when using any of the listed lots, has the potential to impact the interpretation of gentamicin results during therapeutic monitoring at concentrations up to approximately 2.5 g/mL [5.4 mol/L].
FDA-determined cause: Under Investigation by firm
Action taken
On April 19, 2018, the firm sent an Urgent Medical Device Recall letter to customers. The letter identified the affected lots and advised customers to do the following: Inspect stock, discontinue use of and discard the Dimension and Dimension Vista GENT lots listed in Table 1. Please review this letter with your Medical Director. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the regulatory authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens Technical Support Representative.
Details
- Recall number
- Z-2182-2018
- Initiated
- April 19, 2018
- Terminated
- April 24, 2020
- Product code
- LCD – Enzyme Immunoassay, Gentamicin
- Quantity
- 20373
- Distribution
- Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.
- Recalling firm
- Siemens Healthcare Diagnostics, Inc., Newark, DE
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2181-2018 Class II | Dimension Gentamicin Flex reagent cartridge, DM GENT, gentamicin enzyme immunoassay, Lot…The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the… | 04/19/2018Terminated |