Recall Z-2186-2025

Flexicare Medical (Dongguan) Ltd. · initiated July 11, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

BritePro Solo Mini Single-Use Fiber Optic Laryngoscope Handle REF:040-309U

Codes / lots: Lots: 2007xxxxx to 2012xxxxx (July-2020 to Dec-2020)

Reason for recall

Laryngoscope handles may not illuminate as intended.

FDA-determined cause: Nonconforming Material/Component

Action taken

On July 11, 2025, Flexicare Medical Limited issued a Urgent: Medical Device Recall notification to affected consignees via E-Mail. Flexicare ask consignees to take the following actions: 1. Review Inventory for affected product and immediately quarantine. 2. All affected devices must be manually checked using the following procedure within customer notification. 3. All devices which demonstrate a failure to illuminate shall be withdrawn from use, quarantined, and returned. 4. If tested devices illuminate, they can continue to be used. Please ensure you subsequently continue to follow the Pre-Use Checks information as per the IFU. 5. Share this notice with any users of the product within your distribution network. 6. If your firm has further distributed the product, please provide them with the notification and inform them to notify their customers. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to FDA's MedWatch Adverse Event Reporting program . 8. In the event of light failure, the caregiver should revert to the local standard of care, including oxygenation, sourcing a functional laryngoscope, or relying on another method to provide patient ventilation.

Details

Recall number
Z-2186-2025
Initiated
July 11, 2025
Posted by FDA
August 1, 2025
Product code
CCW – Laryngoscope, Rigid
Distribution
US Nationwide distribution in the states of WA, SD, TX, VA, PA, OK, GA, MI, WY, NC, OH.
Recalling firm
Flexicare Medical (Dongguan) Ltd., Dongguan

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2182-2025 Class IBritePro Solo Single-Use Fiber Optic Mini Handle and Blade Ref: 040-02-0110U (Mac 1)…Laryngoscope handles may not illuminate as intended. 07/11/2025Open, Classified
Z-2183-2025 Class IBritePro Solo Single-Use Fiber Optic Handle and Blade Refs: 040-331U (Mac 1), 040-332U…Laryngoscope handles may not illuminate as intended. 07/11/2025Open, Classified
Z-2184-2025 Class IBritePro Solo Mini with Pro Miller Blade Size 1.5, REF:040-02-0415ULaryngoscope handles may not illuminate as intended. 07/11/2025Open, Classified
Z-2185-2025 Class IBritepro Solo Stubby Single-Use Fiber Optic Laryngoscope Handle REF:040-03-0000ULaryngoscope handles may not illuminate as intended. 07/11/2025Open, Classified
Z-2187-2025 Class IBritePro Solo Single-Use Fiber Optic Laryngoscope Handle Refs:040-310AU and 040-310ULaryngoscope handles may not illuminate as intended. 07/11/2025Open, Classified