Recall Z-2188-1991

Hologic, Inc. · initiated June 27, 2008

Terminated

Product

Hologic Oasis Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is indicated for the estimate of bone mineral density, comparison of measured variables obtained from a given scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

Codes / lots: S/N: 11/01/06 B 1/04/08B 11/01/06 A 6/24/05-A 9/25/06-01 9/25/06-02 9/25/06-03 9/25/06-04 9/28/06-01 9/28/06-02 9/28/06-03 9/12/07B 10/02/07B 10/02/07A and 1/04/08A

Reason for recall

Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate.

FDA-determined cause: Software design

Action taken

Hologic notified customers by registered letter titled Urgent: Medical Device Recall, dated 6/27/08, providing instructions to update the software to Version 2.3 and discard version APEX 2.0, 2.1 and 2.2 manuals. If you have questions or require assistance with the software installation, contact Hologic's Help Desk at 1-800-321-4659.

Details

Recall number
Z-2188-1991
Initiated
June 27, 2008
Posted by FDA
September 17, 2008
Terminated
October 25, 2010
Product code
KGI – Densitometer, Bone
Quantity
15 units
Distribution
Nationwide Foreign: Canada Algeria, AE, Australia, Belgium,China,France, Germany, Hungary, Hong Kong, Iceland, Iran, Ireland, Israel, Italy,Japan,Korea, Kuwait, Lithuania, Mexico, Myanmar, Netherlands, Pakistan, Peru, Russia, S. Africa, Saudi Arabia, Singapore, Slovak Rep., Switzerland, Syria, Taiwan, UK
Recalling firm
Hologic, Inc., Bedford, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2184-2008 Hologic Discovery Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is…Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate. 06/27/2008Terminated
Z-2185-2008 Hologic QDR4500 Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is…Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate. 06/27/2008Terminated
Z-2186-2008 Hologic Delphi Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is…Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate. 06/27/2008Terminated
Z-2187-2008 Hologic Explorer Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is…Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate. 06/27/2008Terminated