Recall Z-2202-2026

Becton, Dickinson and Company · initiated April 8, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 831F75/831F75P, TS105F5, 774F75, 777F8, AIQSGF8;

Codes / lots: Model/UDI-DI/Lot(Expiration): 131F7P/00690103000412/65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026),65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026),65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026), 66010507(9/13/2026), 66010508(9/14/2026), 66023074(9/30/2026), 66033088(10/5/2026), 66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026)…

Reason for recall

Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

FDA-determined cause: Process change control

Action taken

On 4/8/2026, mailed correction notices were mailed to customers and distributors who were informed of the following: Consider the following to reduce the potential for hub breakage issues. Minimize manipulation of proximal injectate hub and lumen. Pay frequent attention to the (blue) proximal injectate lumen for leakage or breakage, and exchange with another catheter if the issue is encountered. Modify existing clinical workflows to minimize flexing and stress of (blue) proximal injectate lumen. For example: o Securing the catheter lumens to an IV arm board to mitigate bending/flexing of the lumen(s) o Consider using the stopcock on white hub to allow for intermittent transducer pressure tubing changes without twisting white hub on Swan Ganz (blue) proximal injectate lumen. o Prevent CO-Set syringe and proximal injectate lumen from dangling/hanging without support. o Change to automated cardiac output models to mitigate device stress. Providers and institutions may choose to utilize transthoracic or transesophageal echocardiography in combination with techniques to monitor central venous pressure via a central venous catheter in patients in need of right ventricle monitoring and management. Converting to Acumen IQ / FloTrac sensor can provide continuous minimally invasive hemodynamic monitoring from the arterial line. Consider using ClearSight cuff for continuous non-invasive hemodynamic monitoring. Distributors should forward the correction notice to customers who have purchased the impacted product. Complete and return the acknowledgement form via email to SMB-FCAUSAPM@bd.com Questions can be directed to the firm's Customer Service at (800) 424-3278

Details

Recall number
Z-2202-2026
Initiated
April 8, 2026
Posted by FDA
May 15, 2026
Product code
DYB – Introducer, Catheter
Quantity
589,223
Distribution
Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius…
Recalling firm
Becton, Dickinson and Company, Irvine, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2203-2026 Class IISwan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755PCatheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal… 04/08/2026Open, Classified
Z-2204-2026 Class IISwan-Ganz Pacing Catheter, Models: D200F7;Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal… 04/08/2026Open, Classified