Recall Z-2207-2023

Medtronic Neurosurgery · initiated June 13, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2, 61100; DUREPAIR DURA SUBSTITUTE 3 X 3, 61105; DUREPAIR DURA SUBSTITUTE 1 X 3, 61106; DUREPAIR DURA SUBSTITUTE 4 X 5, 61110; DUREPAIR DURA SUBSTITUTE 1 X 1, 61111; DURA 62100 SUBSTITUTE DUREPAIR 2X2IN NCE, 62100; DURA 62105 SUBSTITUTE DUREPAIR 3X3IN NCE, 62105; DURA 62106 SUBSTITUTE DUREPAIR 1X3IN NCE, 62106; DURA 62110 SUBSTITUTE DUREPAIR 4X5IN NCE, 62110; DURA 62111 SUBSTITUTE DUREPAIR 1X1IN NCE, 62111; DURA 62158 SUBSTITUTE DUREPAIR 5X8IN NCE, 62158.

Codes / lots: REF/Expiration Date On or Prior/UDI-DI/Lots: 61100/March 31, 2025/00643169063808/2006039, 2007020, 2009028, 2010001, 2010016, 2102001, 2102020, 2103001, 2105028, 2106001, 2108011, 2108017, 2109022, 2110009, 2110030, 2111023, 2111045, 2112026, 2203020, 2204016, 2204035, 2207014, 2207046; 61105/February 28, 2025/00643169063815/2009060, 2011017, 2102011, 2102027, 2102047, 2103002, 2104042, 2105029, 2108001, 2108012, 2108016, 2109023, 2110010, 2111021, 2111024, 2111046, 2112027, 2202034, 2203034, 2204027, 2206008, 2207015, 2207033; 61106/December 31, 2024/00643169063822/2010017, 2103003, 2105030, 2105038, 2108013, 2108018, 2109004, 2110011, 2110031, 2111025, 2202035, 220302; 61110/March 31, 2025/00643169063839/2006039, 2010018, 2101014, 2102002, 2102028, 2103004, 2105031, 2108019, 2109024, 2110012, 2111026, 2201005, 2202036, 2204025, 2204040; 61111/April 30, 2025/00643169063846/2008039, 2010019, 2102021, 2103005, 2105032, 2106004, 2108014, 2110013, 2111027, 2202037, 2203022, 2203028, 2207016; 62100/January 31, 2025/00643169063853/2007021, 2008040, 2009026, 2009061, 2010014, 2011018, 2012001, 2101015, 2101020, 2102012, 2102022, 2102048, 2103019, 2104043, 2105033, 2106005, 2108020…

Reason for recall

Dura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may present clinically with signs and symptoms of an acute inflammatory process, comparable to infection.

FDA-determined cause: Process change control

Action taken

On 6/13/23 recall notices were mailed to customers and distributors and they were asked to do the following: 1) Return affected product to the recalling firm. 2) Complete and return the Customer Confirmation Form to neuro.quality@medtronic.com 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Customers with questions can contact Customer Service at 800-874-5797 Option 1.

Details

Recall number
Z-2207-2023
Initiated
June 13, 2023
Posted by FDA
July 20, 2023
Product code
GXQ – Dura Substitute
Quantity
65029
Distribution
Worldwide - US Nationwide distribution including in the states of PR, TX, WA, HI, MI, GA, AL, IL, IN, IA, MO, KY, NY, CA, AZ, PA, UT, ME, MD, MN, FL, AK, NE, VA, OR, WI, NC, LA, SD, TN, MS, DE, NJ, ID, SC, OH, MA, DC, NM, WV, NH, CT, NV, OK, AR, KS, MT, RI, CO, ND, WY, VT and the countries of Barbados, Cayman Islands, Honduras, Belgium, Switzerland, Italy, United Kingdom, Canary Islands…
Recalling firm
Medtronic Neurosurgery, Irvine, CA

Official FDA recall record