Recall Z-2207-2026
Draeger, Inc. · initiated May 7, 2026
Class I – dangerous or defective products that could cause serious health problems or death
Product
Atlan A350. Model Number: 8211500. anesthesia workstation
Codes / lots: Model Number: 8211500. UDI-DI Number: 04048675556176.Serial Numbers: ASSL-0347, ASSL-0348, ASSL-0351, ASSH-0220, ASSH-0221, ASSH-0224, ASSH-0225, ASSH-0228, ASSH-0229, ASSH-0230, ASSH-0234, ASSH-0235, ASTA-0086, ASSL-0349, ASSL-0350, ASSL-0352, ASSL-0353, ASSH-0222, ASSH-0223, ASSH-0226, ASSH-0227, ASSH-0231, ASSH-0232, ASSH-0233, ASSH-0236, ASSH-0237, ASSH-0238, ASSN-0185, ASSN-0186, ASSN-0187, ASSN-0188, ASSN-0189, ASSN-0190, ASTB-0037, ASTB-0040, ASTA-0087, ASTA-0088, ASSM-0130, ASSM-0131, ASSM-0132.
Reason for recall
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
FDA-determined cause: Process change control
Action taken
Draeger notified consignees of the recall update and expansion on about 05/07/2026 via emailed letter titled Urgent Medical Device Correction Update. Consignees were made aware that due to continuous monitoring, Draeger has determined that an additional number of Atlan devices are affected by the issue and they are replacing ventilator motors. Consignees were instructed that a Draeger Service Representative will contact consignees to arrange a date for the ventilator motor assembly replacement, and until the corrective action is implemented, the device may continue to be used with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient manually as needed. Consignees were also instructed to complete and return the provided Medical Device Correction Return Response Acknowledgment and Receipt Form and if products have been made available to third parties, please forward this Urgent Medical Device Correction notice to them.
Details
- Recall number
- Z-2207-2026
- Initiated
- May 7, 2026
- Posted by FDA
- June 4, 2026
- Product code
- BSZ – Gas-Machine, Anesthesia
- Quantity
- 3,515 units
- Distribution
- US distribution in WI, AR, MN, WI, CO, MA, PA, AZ, FL, NY, SD, ND. International distribution to: Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chili, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Gabon, Germany, Ghana, Greece, Greenland, Hong Kong, Hungary, India…
- Recalling firm
- Draeger, Inc., Telford, PA