Recall Z-2210-2019

Bard Peripheral Vascular Inc · initiated May 2, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

EnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton, single use, Rx, catalog/REF numbers 10GV Upright (Vertical) - REF ECP0110GV. Product Usage: Indicated for use in acquiring tissue for diagnosing breast abnormalities.

Codes / lots: Lot numbers VTCT0273, VTCT0308, VTCT0309, VTCU0351, VTCU0352, VTCW0451, VTCW0452, VTCW0453, VTCX0485, VTCX0486, VTCX0487, VTCY0548, VTCY0549, VTCY0550, VTCZ0642, VTCZ0643, VTCZ0644, and VTCZ0645.

Reason for recall

There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product.

FDA-determined cause: Component change control

Action taken

The recalling firm issued letters dated 4/19/2019 via FedEx on 4/19/2019 with proof of delivery to the U.S. customers. The letters sent to foreign countries vary depending on the format required for the country.

Details

Recall number
Z-2210-2019
Initiated
May 2, 2019
Terminated
March 3, 2022
Product code
KNW – Instrument, Biopsy
Quantity
29,918 units
Distribution
Worldwide distribution - US Nationwide distribution, including Puerto Rico. There was government and foreign distribution and no military distribution. Foreign distribution was made to Canada, Algeria, Australia, Austria, Belgium, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, India…
Recalling firm
Bard Peripheral Vascular Inc, Tempe, AZ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2209-2019 Class IIEnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton…There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples… 05/02/2019Terminated
Z-2211-2019 Class IIEnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton…There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples… 05/02/2019Terminated
Z-2212-2019 Class IIEnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton…There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples… 05/02/2019Terminated
Z-2213-2019 Class IIEnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton…There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples… 05/02/2019Terminated
Z-2214-2019 Class IIEnCor Breast Biopsy Probes, packaged individually in sterile pouches, 5 pouches/carton…There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples… 05/02/2019Terminated