Recall Z-2217-2019
Medical Components, Inc dba MedComp · initiated March 13, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
14F x 30CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061425 Product Usage: An aid for introducing a hemodialysis catheter
Codes / lots: MCAM240 MCBL480 MCBS870 MCCF480 MCCJ320 MCCW690 MCCZ500 MCFG830 MCFR160 MCFS760 MCFZ360 MKAJ330 MKAR130 MKBL500 MKBX060 MMAB730
Reason for recall
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
FDA-determined cause: Labeling Change Control
Action taken
On March 13, 2019, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed that the kits contain a 15F "non-valved" peelable introducer despite the label indicating that the kit is packaged with a "valved" introducer. Customers were asked to do the following: 1. Immediately examine your inventory and quarantine product subject to recall. 2. If you have further distributed this product, please identify your customers and immediately notify them of the recall. 3. Contact your Medical Components, Inc. customer service representative for a Returned Goods Authorization (RGA) number if you have product to return. The phone number to call is 215-256-4201. 4.Complete and return the Medical Device Recall Return Response Form. The form can be returned by fax (215-256-9191) or email (recalls@medcompnet.com). Contact Information for questions: Monday through Friday, 8:00 AM to 4:00 PM, Eastern Standard Time. Phone: 215-256-4201; Email: recalls@medcompnet.com
Details
- Recall number
- Z-2217-2019
- Initiated
- March 13, 2019
- Terminated
- March 5, 2024
- Product code
- LFJ – Catheter, Subclavian
- Quantity
- 6830
- Distribution
- US Nationwide Distribution CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
- Recalling firm
- Medical Components, Inc dba MedComp, Harleysville, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2215-2019 Class II | 14F x 20CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No.…Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a… | 03/13/2019Terminated |
| Z-2216-2019 Class II | 14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No.…Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a… | 03/13/2019Terminated |