Recall Z-2220-2010
Insight Instruments, Inc. · initiated May 12, 2010
Product
CORNEA COAT Hydroxypropylmethylcellulose 1.5ml Syringe. STERILE, Rx ONLY. Manufactured by: Insight Instruments, Inc., 2580 SE Willoughby Blvd. Stuart, FL 34994 USA. REF 60010.000, QTY 14 - This package contains 14 individual packs, each containing 1 sterile, single use ocular irrigation device. For external use only! Used as an ocular irrigation device.
Codes / lots: Product Item Lot # Cornea Coat Lot # Exp Date Cornea Coat M408080708 M408080708 2010-06 Cornea Coat M414420808 M414420808 2010-06 Cornea Coat M415940808 M415940808 2010-06 Cornea Coat M421610908 M421610908 2010-08 Cornea Coat M427661108 M427661108 2010-09 Cornea Coat M434311108 M434311108 2010-09 Cornea Coat M439690109 M439690109 2010-10 Cornea Coat M44744 M44744 2010-12
Reason for recall
Insight Instruments, Inc., Stuart, FL is recalling CORNEA COAT 2.5% Hydroxypropylmethylcellulose solution in 1.5 ml Syringe due to a potential weakness of the peel pouch seals. Products affected are Item Numbers 600100.000 sold individually and 40201.000 thru 10206.000 sold as components of BIOM Pack.
FDA-determined cause: Packaging
Action taken
All consignees were notified of the recall by letter sent via first class mail. Letter contained a Return Response Form.
Details
- Recall number
- Z-2220-2010
- Initiated
- May 12, 2010
- Posted by FDA
- August 13, 2010
- Terminated
- February 16, 2012
- Product code
- KYG – Device, Irrigation, Ocular Surgery
- Quantity
- 16,128
- Distribution
- Worldwide Distribution -- USA, including Puerto Rico, and Frankfurt, Germany.
- Recalling firm
- Insight Instruments, Inc., Stuart, FL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2221-2010 | BIOM PACK, Contents Individually Sterile. QTY 5 EA. Each Pack Contains: 1 EA. Microscope…Insight Instruments, Inc., Stuart, FL is recalling CORNEA COAT 2.5% Hydroxypropylmethylcellulose solution in… | 05/12/2010Terminated |