Recall Z-2224-2026

Fresenius Kabi USA, LLC · initiated May 6, 2026

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Codes / lots: Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:

Reason for recall

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

FDA-determined cause: Device Design

Action taken

On May 6, 2026 URGENT - Medical Device Field Correction letters were sent to customers. Required Actions for Healthcare Providers 1. If a unit has been dropped or severely jarred, remove it from service, even if no damage is visible. 2. Once you have removed the pump from service, or if you have questions about whether to do so after a drop, please contact Fresenius Kabi customer support at Ivenix_support@fresenius-kabi.com or (855) 354-6387. 3. Post enclosed Safety Tip Sheet, Proper Handling of the Ivenix Large Volume Pump (LVP) notice within your facilities to ensure that all ancillary staff are informed of this notification and trained on the action to remove any dropped or severely jarred devices from use. 4. Notify other users if applicable. If your facility distributes or transfers devices to other locations, notify all anticipated users of this correction.

Details

Recall number
Z-2224-2026
Initiated
May 6, 2026
Posted by FDA
May 29, 2026
Product code
FRN – Pump, Infusion
Quantity
18,444
Distribution
Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
Recalling firm
Fresenius Kabi USA, LLC, North Andover, MA

Official FDA recall record