Recall Z-2230-2021
Shent USA, Inc. · initiated January 12, 2021
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used through the working channel of a GI endoscope
Codes / lots: Lot Code: D20177226-19
Reason for recall
Mislabeled: Devices labeled as catalog #7208 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm) may actually be catalog #7226 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 8 mm x 230 cm) and Devices labeled as catalog #7226, (Grasper/Retriever, Alligator Jaw, Rat Tooth, 2.4 mm x 8 mm x 230 cm) may actually be catalog #7208, Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm).
FDA-determined cause: Packaging process control
Action taken
Telemed issued Urgent Medical Device Recall letter on 1/12/21. Letter states reason for recall, health risk and action to take: Please respond to this recall notification (email response is acceptable) to let us know whether or not you have any of the recalled product in your inventory. If you are in possession of any recalled devices, please provide the quantity, catalog number and lot number. All affected devices must be returned to TeleMed; upon receiving your response to this notification, TeleMed Customer Service will provide you with an RGA number for product return. If you have any questions regarding this medical device recall, please contact: prakash@telemedsystems.com. 508-723-6609; Customerservice@telemedsystems.com 978-567-9033
Details
- Recall number
- Z-2230-2021
- Initiated
- January 12, 2021
- Product code
- OCZ – Endoscopic Grasping/Cutting Instrument, Non-Powered
- Quantity
- 85 units OUS
- Distribution
- US Nationwide Distribution - CA, CO, MI Foreign: Australia
- Recalling firm
- Shent USA, Inc., Hudson, MA