Recall Z-2236-2014

Nanosphere, Inc. · initiated July 9, 2014

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Verigene BC-GN Test Kit (Catalog number 20-005-021) includes: 1) 20 Verigene BC-GN Test Cartridges. Each Test Cartridge comes preloaded with all required reaction solutions, including wash solutions, oligonucleotide probe solution and signal amplification solutions required to generate a test result. BC-GN is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, is not used to monitor these infections.

Codes / lots: VERIGENE BC-GN TEST KIT (Catalog Number: 20-005-021). TEST CARTRIDGES: Part Number: 20-006-021, Lot Number: 042914021B, Expiration Date: 10/26/14

Reason for recall

Nanosphere Inc. has recently determined through four customer complaints that a small number of Verigene Gram-Negative Blood Culture Nucleic Acid Test (BC-GN) Extraction Trays in Lot 042914021B have low levels or are missing Magnetic Microparticles (MMPs) required for DNA Extraction.

FDA-determined cause: Under Investigation by firm

Action taken

Nanosphere sent an Urgent Product Recall letter dated July 9, 2014 to all affected customers. All customers were initially notified of the recall by telephone on July 9, 2014. The customers were instructed to quarantine and return all affected product to Nanosphere. The letter included instructions: 1) on how to return the recalled products, and 2) to review any test results that were generated while using the recalled products to ensure that there has been no negative impact on patient care. Customers with questions were instructed to contact Nanosphere at 888-837-4436, option 2 or via e-mail at Technicalsupport@nanosphere.us.

Details

Recall number
Z-2236-2014
Initiated
July 9, 2014
Posted by FDA
August 14, 2014
Terminated
September 1, 2015
Product code
PEN – Gram-Negative Bacteria And Associated Resistance Markers
Quantity
1045 extraction trays (840 affected by recall)
Distribution
Nationwide Distribution including CA, CO, IA, MT, OH, OK, OR, TX, and UT
Recalling firm
Nanosphere, Inc., Northbrook, IL

Official FDA recall record