Recall Z-2244-2016

Bayer Healthcare · initiated June 9, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Angiographic Injector and Syringe, MEDRAD MRXperion MR Injection System, MEDRAD MRXperion Sterile Disposable MRI Kit Usage: The MEDRAD MRXperion MR Injection System is a syringe based fluid delivery system indicated for delivery of contrast media and saline during MR applications. It is intended to be used for the specific purpose of injecting intravenous MR contrast media and saline into the human vascular system for diagnostic studies in magnetic resonance imaging (MRI) applications with MRI scanners that have a magnetic field strength between 0.7 and 3.0 Tesla. Only trained healthcare professionals are intended to operate this device.

Codes / lots: Catalog numbers: MRXP 200 Serial numbers: 100009, 100011, 100019, 100020, 100022, 100026, 100027, 100028, 100036, 100047, 100065, 100066, 100069, 100074, 100075, 100085, 100090, 100091, 100096, 100097, 100098, 100099, 100100, 100101, 100103, 100104, 100106, 100107, 100108, 100109, 100110, 100111, 100112, 100113, 100114, 100118, 100119, 100129, 100152, 100155, 100161, 100166

Reason for recall

Bayer Healthcare is initiating this recall due to complaints that were received from customer sites describing a 4205 error message when the injector is used with a 3T scanner.

FDA-determined cause: Device Design

Action taken

Bayer sent an Urgent Medical Device Field Corrective Action letter dated June 8, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to take the following steps should a 4205 error message is encountered. For questions contact Bayer Customer Support Team at 1-800-633-7237.

Details

Recall number
Z-2244-2016
Initiated
June 9, 2016
Terminated
December 5, 2016
Product code
DXT – Injector And Syringe, Angiographic
Quantity
42 units
Distribution
Worldwide and the US state of NEW YORK
Recalling firm
Bayer Healthcare, Indianola, PA

Official FDA recall record