Recall Z-2245-2021

BioFire Diagnostics, LLC · initiated July 20, 2021

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127

Codes / lots: Part No: RFIT-ASY-0126 / UDI: 00815381020086; Part No: RFIT-ASY-0127 / UDI: 00815381020093; Kit Lot# 0868221; Pouch Lot# 194Z21; Batch 003

Reason for recall

Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.

FDA-determined cause: Process control

Action taken

On 07/20/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customer informing them that an issue has been identified in the manufacturing process which might contribute to elevated rates of false negative results while using the BCID Panel. Actions to be taken by customers: 1) Immediately examine their inventory for product identified in this recall, 2) Discontinue use and discard any remaining product in their possession that is subject to this recall. The Recalling Firm will replace the product at no charge in accordance with its standard limited warranty. 3) Complete the accompanying Field Action Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to the Recalling Firm so that it may acknowledge their receipt of this notification. 4) If they may have further distributed the affected product, identify any recipients and notify them at once of this product recall. For product replacement, as well as any other questions or concerns, contact customer support department at Support@biofiredx.com or via telephone at 1-800-735-6544 and selecting option 5 for Product Technical Support.

Details

Recall number
Z-2245-2021
Initiated
July 20, 2021
Terminated
December 16, 2022
Product code
PEN – Gram-Negative Bacteria And Associated Resistance Markers
Quantity
29 kits
Distribution
U.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None
Recalling firm
BioFire Diagnostics, LLC, Salt Lake City, UT

Official FDA recall record