Recall Z-2247-2023

Johnson & Johnson Surgical Vision, Inc. · initiated May 30, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Codes / lots: All lots within their expiration date as of 5/29/2023. Part Number/UDI-DI/Lot: VRT-AI/05050474700901/60298931, 60302489, 60302939, 60302941, 60302953, 60302954, 60304170, 60304175, 60304176, 60305660, 60305661, 60305662, 60305665, 60306934, 60306935, 60306936, 60306940, 60306942, 60308193, 60309850, 60309844, 60309845, 60309853, 60314642, 60314676, 60314677, 60316112, 60351692, 60372490, 60372491, 60374175, 60374945, 60376930, 60378840, 60378842, 60378844, 60400795, 60401984, 60413115, 60413116, 60413117, 60414191, 60415077, 60415078, 60416304, 60420003, 60421952, 60425389, 60426314, 60426315, 60428217, 60428216, 60429442, 60433653, 60433654, 60433655, 60435921, 60435922, 60435923, 60437988, 60437989, 60440239, 60440240, 60442450, 60442451, 60442452, 60442453, 60442454, 60444535, 60446059, 60448847, 60448856, 60448857, 60376931, 60379976, 60429443. VRT-AF/05050474700895/60304172, 60304173, 60304174, 60305657, 60305658, 60305659, 60306930, 60306931, 60306932, 60306933, 60308189, 60308190, 60309848, 60309847, 60314640, 60314641, 60314674, 60314675, 60378839, 60420004, 60446054, 60446055

Reason for recall

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

FDA-determined cause: Process control

Action taken

On 5/30/23, recall notices were mailed to customers who were asked to do the following: 1) Discontinue use and return affected devices to the recalling firm. 2) Share this notice with anyone within your organization that needs to be informed and to any organization where the potentially affected products have been transferred. 3) Complete and return the customer reply form to RegCompliOne@its.jnj.com In addition, distributors were asked to distribute the recall notice to their customers. Questions, adverse events, and complaints can be directed to the firm at 1-877-266-4543 Option 2.

Details

Recall number
Z-2247-2023
Initiated
May 30, 2023
Posted by FDA
July 28, 2023
Product code
HQC – Unit, Phacofragmentation
Quantity
148,941
Distribution
Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong…
Recalling firm
Johnson & Johnson Surgical Vision, Inc., Irvine, CA

Official FDA recall record