Recall Z-2248-2020

Radiometer Medical ApS · initiated March 17, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

safeCLINITUBES Blood Sampler;REF 942-898;D957P-70-70x1

Codes / lots: Lot numbers: R1431 - R1495

Reason for recall

The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost samples.

FDA-determined cause: Process control

Action taken

The method of notification of the recall was a customer letter sent to all consignees via email and a hard copy sent via FedEx 2nd day with tracking and proof of delivery; date of issue was 3/17/20. Customers were instructed to do the following with the recalled product: -Cease using the Blood Samplers from the affected lots. -Check inventory of Blood Samplers from the affected lots. -Check for Blood Samplers from the affected lots distributed in the institution. -Collect any Blood Samplers from the affected lots. -Complete the Recall Response Form (last page of the customer advisory letter) and return it to the representative. -Discard the collected Blood Samplers of the affected lots. For any questions contact at 1-800-736-0600 option 1, Monday-Friday, 9am EST to 7pm EST.

Details

Recall number
Z-2248-2020
Initiated
March 17, 2020
Terminated
July 28, 2022
Product code
GIO – Tube, Collection, Capillary Blood
Quantity
269 units
Distribution
U.S.: AK, AL, AR, CA, CO, CT, FL,GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, and WI. OUS: Australia (New Zealand), Austria, Brazil Canada, China, Czech Republic (Slovakia), Denmark, Dubai (United Arab Emirates), Egypt, Finland, France, Germany, Hong Kong, Hungary, Iceland, India, Ireland, Italy, Japan, Kazakhstan…
Recalling firm
Radiometer Medical ApS, Bronshoj

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2247-2020 Class IIsafeCLINITUBES Blood Sampler;REF 942-892;D957P-70-100x1The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost… 03/17/2020Terminated
Z-2249-2020 Class IIsafeCLINITUBES Blood Sampler;REF 942-969;D957P-70-45x1The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost… 03/17/2020Terminated