Recall Z-2251-2025

AVID Medical, Inc. · initiated June 11, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

GYN ROBOTIC PACK. Medical convenience kit.

Codes / lots: Model No. NAMC474-01; UDI: 10809160375044; Kit Lot No. 1620573, 1620572, 1621883, 1621884, 1628673.

Reason for recall

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

FDA-determined cause: Under Investigation by firm

Action taken

Consignees were sent an URGENT: MEDICAL DEVICE FIELD CORRECTION notice, dated 6/11/25, via email. The notification instructs consignees to not use the CLEARIFY Visualization System (component #5303907) and discard. All other components in the kit may be continued to be used. Consignees are to add warning labels to affected procedure trays to ensure the affected component is discarded. The provided Recall Response Form is to be completed and returned to Owens&Minor. Any quality problems or adverse events associated with recalled devices may be reported via email to complaints@owens-minor.com

Details

Recall number
Z-2251-2025
Initiated
June 11, 2025
Posted by FDA
August 4, 2025
Product code
OKV – Obstetrical Kit
Quantity
45 units
Distribution
Domestic: CA, IL, MD, OH, PA, TX, WA.
Recalling firm
AVID Medical, Inc., Toano, VA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2240-2025 Class IILAPAROSCOPIC TRAY. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2241-2025 Class IICARDIAC ROBOT PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2242-2025 Class IICT DAVINCI. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2243-2025 Class IIDA VINCI PACK URO/PROSTATE. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2244-2025 Class IIDAVINCI TRUCUSTOM BUNDLE. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2245-2025 Class IIDAVINCI UROLOGY. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2246-2025 Class IIGEN LAPAROSCOPY PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2247-2025 Class IIGENERAL LAPAROSCOPY TRAY. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2248-2025 Class IIGENERAL ROBOTIC PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2249-2025 Class IIGS ENDOSCOPY PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2250-2025 Class IIGYN LAPAROSCOPY PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2252-2025 Class IIHVI CARDIAC ROBOT PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2253-2025 Class IILAP CHOLE PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2254-2025 Class IILAPAROSCOPIC GYN. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2255-2025 Class IILAPAROSCOPIC PROSTATECTOMY. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2256-2025 Class IILAPAROSCOPY PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2257-2025 Class IILAVH PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2258-2025 Class IIMAJOR THORACOSCOPY BASIN. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2259-2025 Class IIPACK GENERAL ROBOTIC. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2260-2025 Class IIPARTIAL NEPHRECTOMY - ROBOTIC. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2261-2025 Class IIPARTIAL NEPHRECTOMY PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2262-2025 Class IIROBOTIC PROSTATECTOMY PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified
Z-2263-2025 Class IIROBOTIC URO/GYN PACK. Medical convenience kit.CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration. 06/11/2025Open, Classified