Recall Z-2263-2016

The Metrix Company · initiated June 23, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Dual Chamber Container fill with gravity/attached transfer sets, 250 mL / 1250 mL, 300 mL / 1700 mL, 500 mL / 2500 mL, 500 mL / 3500 mL

Codes / lots: All lots are being recalled. Lot code consists of the five digit product code followed by a dash followed by a unique five character numeric or alphanumeric sequence. Affected product codes for this product: 66617, 66622, 66632, 66642, 69616, 69621, 69631, and 69641

Reason for recall

An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.

FDA-determined cause: Device Design

Action taken

On 6/23/2016, the recalling firm sent written recall notices to their distributors and end users notifying them of the recall. Customers and consignees were instructed to not use the devices, locate and quarantine all affected devices, and to complete the response form.

Details

Recall number
Z-2263-2016
Initiated
June 23, 2016
Terminated
April 11, 2023
Product code
NEP – System/Device, Pharmacy Compounding
Quantity
18,474 case
Distribution
Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United Kingdom
Recalling firm
The Metrix Company, Dubuque, IA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2257-2016 Class IIAUTOCOMP6 XPS High Speed Compounder REF 58810An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated
Z-2258-2016 Class IIAUTOCOMP6 XP High Speed Compounder REF 58810An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated
Z-2259-2016 Class IIAUTOCOMP6 High Speed CompounderAn issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated
Z-2260-2016 Class IIContainer fill and transfer sets, 250 mL, 500 mL, 1000 mL, 2000 mL, 3000 mL, or 4000 mL.An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated
Z-2261-2016 Class IIContainer fill and transfer sets, automatic fill compatible, 250 mL, 500 mL, 1000 mL…An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated
Z-2262-2016 Class IIDual Chamber Container fill and transfer sets, 250 mL / 1250 mL or 500 mL / 2500 mLAn issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated
Z-2264-2016 Class IIContainer fill and transfer sets, 250 mL and 500 mL, . .An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the… 06/23/2016Terminated