Recall Z-2265-2012

Canon Inc. · initiated November 1, 2010

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

Codes / lots: Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA.

Reason for recall

Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.

FDA-determined cause: Nonconforming Material/Component

Action taken

Canon Inc. notified sales companies including Canon USA on 9/15/2010 with Service Bulletin 10-017, and on 8/25/2011 with Service Bulletin 11-011, and on 1/20/2011 with Service Bulletin 11-001. Sales representatives were to visit customers and check and replace the screws.

Details

Recall number
Z-2265-2012
Initiated
November 1, 2010
Posted by FDA
August 24, 2012
Terminated
January 7, 2014
Product code
MQB – Solid State X-Ray Imager (Flat Panel/Digital Imager)
Quantity
468 units
Distribution
Nationwide Distribution (USA)
Recalling firm
Canon Inc., Ohta-ku, Tokyo

Official FDA recall record