Recall Z-2265-2019
Degania Silicone, Ltd. · initiated June 28, 2019
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Codes / lots: Lot Codes: P1851056 P1867059
Reason for recall
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
FDA-determined cause: Process control
Action taken
Degania Silicone Ltd. notified accounts by letter dated June 28, 2019 describing the problem, health risk and action to take: Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list.2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Promptly return the completed Recall Response form to Degania Silicone by e-mail: kkayam@qco.net. 4. Disposition of the recalled product: - the preferred option for us is to scrap any recalled product at your facility. In this case we will require official Destruction Confirmation letter containing details of the scrapped product, signed by your company (the draft will be provided by request). - if the above option is not acceptable, we would ask to return any recalled product to our Rhode Island facility. Questions contact Degania Silicone: Degania Silicone Ltd. Degania Bet, 1513000, Israel Katia Kayam, Customer complaints Officer Tel: +972-529532059 E-mail: Kkayam@qco.net
Details
- Recall number
- Z-2265-2019
- Initiated
- June 28, 2019
- Terminated
- October 22, 2020
- Product code
- GBX – Catheter, Irrigation
- Quantity
- 2160
- Distribution
- Nationwide Distribution
- Recalling firm
- Degania Silicone, Ltd.
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2260-2019 Class II | Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |
| Z-2261-2019 Class II | Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes Cardinal Health PN:…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |
| Z-2262-2019 Class II | Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |
| Z-2263-2019 Class II | Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |
| Z-2264-2019 Class II | Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes, with trocar Cardinal…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |
| Z-2266-2019 Class II | Medline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |
| Z-2267-2019 Class II | Smith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… | 06/28/2019Terminated |