Recall Z-2265-2019

Degania Silicone, Ltd. · initiated June 28, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Codes / lots: Lot Codes: P1851056 P1867059

Reason for recall

15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body

FDA-determined cause: Process control

Action taken

Degania Silicone Ltd. notified accounts by letter dated June 28, 2019 describing the problem, health risk and action to take: Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list.2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Promptly return the completed Recall Response form to Degania Silicone by e-mail: kkayam@qco.net. 4. Disposition of the recalled product: - the preferred option for us is to scrap any recalled product at your facility. In this case we will require official Destruction Confirmation letter containing details of the scrapped product, signed by your company (the draft will be provided by request). - if the above option is not acceptable, we would ask to return any recalled product to our Rhode Island facility. Questions contact Degania Silicone: Degania Silicone Ltd. Degania Bet, 1513000, Israel Katia Kayam, Customer complaints Officer Tel: +972-529532059 E-mail: Kkayam@qco.net

Details

Recall number
Z-2265-2019
Initiated
June 28, 2019
Terminated
October 22, 2020
Product code
GBX – Catheter, Irrigation
Quantity
2160
Distribution
Nationwide Distribution
Recalling firm
Degania Silicone, Ltd.

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2260-2019 Class IICardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated
Z-2261-2019 Class IICardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes Cardinal Health PN:…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated
Z-2262-2019 Class IICardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated
Z-2263-2019 Class IICardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated
Z-2264-2019 Class IICardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes, with trocar Cardinal…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated
Z-2266-2019 Class IIMedline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated
Z-2267-2019 Class IISmith & Nephew Negative Pressure Wound Therapy 15FR Channel Wound Drain PN: 02872 An…15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection… 06/28/2019Terminated