Recall Z-2268-2010

Hill-Rom, Inc. · initiated April 13, 2009

Terminated

Product

Hill-Rom 1000 bed model P1160.

Codes / lots: K029AP6787, K029AP6788, K029AP6790, J317AP5267, J206AP2503, J206AP2506, J206AP2511, J206AP2513, J213AP2581, J220AP2773, J220AP2775, J220AP2777, J220AP2778, J224AP2904, J224AP2906, J224AP2907, J224AP2908, J220AP2770, J224AP2909, J224AP2916, J227AP3052, J227AP3053, K027AP6733, K027AP6735, K027AP6736, K027AP6738, J220AP2772, J220AP2774, J224AP2905, J274AP4108, J274AP4113, J274AP4114, J274AP4134, J274AP4135, J274AP4136, J274AP4102, J350AP5889, J350AP5890, J350AP5891, J350AP5892, J350AP5893, J350AP5894, J350AP5895, J350AP5896, J350AP5897, J350AP5899, J350AP5900, J350AP5901, J350AP5902, J350AP5903, J350AP5904, J350AP5905, J350AP5906, J350AP5907, J350AP5908, J350AP5909, J350AP5998, J352AP5988, J352AP5989, J352AP5990, J352AP5991, J352AP5992, J352AP5993, J352AP5994, J352AP5995, J352AP5996, J352AP5997, J352AP5998, J352AP5999, J352AP6000, J352AP6001, J352AP6002, J352AP6003, J352AP6004, J352AP6005, J352AP6006, J352AP6007, J352AP6008, J352AP6009, J352AP6010, J352AP6011, J352AP6012, K020AP6555, K020AP6556, K020AP6557, K020AP6558, K020AP6559, K020AP6560, J354AP6044, J354AP6045, J354AP6046, J354AP6047, J354AP6048, J354AP6049, J354AP6050, J354AP6051, J354AP6052, J354AP6053, J354AP6054…

Reason for recall

Pinch hazard. Potential for shearing action between the side rail arm and the bracket used to secure the side rail to the bed.

FDA-determined cause: Device Design

Action taken

Hill-Rom sent a customer notification letter dated 4/9/2009. A second notice was sent to all customers on 7/9/2009. The letter included instructions and a Customer Response Form, along with the required parts. Customers were instructed to do the correction on their affected beds and return the completed customer response form within 45 days.

Details

Recall number
Z-2268-2010
Initiated
April 13, 2009
Posted by FDA
August 24, 2010
Terminated
June 6, 2011
Product code
FNL – Bed, Ac-Powered Adjustable Hospital
Quantity
39,971 both products
Distribution
Worldwide distribution: USA, Afghanistan, Algeria, American Samoa, Argentina, Australia, Bermuda, Brazil, Canada, Chili, China, Colombia, Costa Rica, Cuba, Ecuador, Egypt, Hong Kong, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Kyrgyzstan, Lebanon, Libya, Malaysia, Mexico, Oman, Panama, Paraguay, Philippines, Puerto Rico, Qatar, Saudi Arabia, Singapore, South Korea, Syria, Taiwan…
Recalling firm
Hill-Rom, Inc., Batesville, IN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2267-2010 Hill-Rom Care Assist Bed, Model P1170.Pinch hazard. Potential for shearing action between the side rail arm and the bracket used to secure the… 04/13/2009Terminated